Toshiba TSX-30IC AQUlLlON ONE System, Computed Tomography X-Ray. This system supports Whole body scanning.

Published April 17, 2013 · Recall initiated: November 14, 2012

Summary

FDA records medical device Class II recall Z-1068-2013, published April 17, 2013 and initiated November 14, 2012. Recalling firm: Toshiba American Medical Systems Inc; Status: Terminated; Reported quantity: 1 unit; Lot: Serial No: 1CB1262006. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Due to a problem with the control software of the x-ray high-voltage generator, if scanning is performed with a tube current less than 30mA, the control software may detect an error and perform error processing even though the x-ray output is normal.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Toshiba American Medical Systems Inc
Distributor
Distributed in the state of MD.
Lot
Serial No: 1CB1262006
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Toshiba TSX-30IC AQUlLlON ONE System, Computed Tomography X-Ray. This system supports Whole body scanning. recalled?

Yes. FDA published recall Z-1068-2013 on April 17, 2013. Reported hazard: Due to a problem with the control software of the x-ray high-voltage generator, if scanning is performed with a tube current less than 30mA, the control software may detect an error and perform error processing even though the x-ray output is normal.

What should I do if I have this product?

Voluntary: Firm initiated · Letter