Toshiba TSX-30IC AQUlLlON ONE System, Computed Tomography X-Ray. This system supports Whole body scanning.

Published April 17, 2013 · Recall initiated: November 14, 2012

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Due to a problem with the control software of the x-ray high-voltage generator, if scanning is performed with a tube current less than 30mA, the control software may detect an error and perform error processing even though the x-ray output is normal.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Toshiba American Medical Systems Inc
Distributor
Distributed in the state of MD.
Lot
Serial No: 1CB1262006
Risk classification
Class II
Area (FDA)
device
Country
United States