Toshiba TSX-30IC AQUlLlON ONE System, Computed Tomography X-Ray. This system supports Whole body scanning.
Published April 17, 2013 · Recall initiated: November 14, 2012
Summary
FDA records medical device Class II recall Z-1068-2013, published April 17, 2013 and initiated November 14, 2012. Recalling firm: Toshiba American Medical Systems Inc; Status: Terminated; Reported quantity: 1 unit; Lot: Serial No: 1CB1262006. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Due to a problem with the control software of the x-ray high-voltage generator, if scanning is performed with a tube current less than 30mA, the control software may detect an error and perform error processing even though the x-ray output is normal.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Toshiba American Medical Systems Inc
- Distributor
- Distributed in the state of MD.
- Lot
- Serial No: 1CB1262006
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAINFX-8000C Fluoroscopic X-Ray SystemsMarch 14, 2018
- FDAAPLIO 500 TUS-A500 diagnostic ultrasound systemJuly 19, 2017
- FDAXARIO 200 TUS-X200 diagnostic ultrasound systemJuly 19, 2017
- FDAToshiba Kalare DREX-KL80 X-ray generatorMarch 22, 2017
- FDATSX-IOI: AQUILION 32/64; System, X-Ray, Tomography, Computed. Designed to produce cross-sectional images of aApril 17, 2013
- FDAIMPAX CV 7.8 SU3 - OCR Service. The IMPAX CV Reporting component facilitates the quick , effective creation ofApril 17, 2013
Questions about this recall
Was Toshiba TSX-30IC AQUlLlON ONE System, Computed Tomography X-Ray. This system supports Whole body scanning. recalled?
Yes. FDA published recall Z-1068-2013 on April 17, 2013. Reported hazard: Due to a problem with the control software of the x-ray high-voltage generator, if scanning is performed with a tube current less than 30mA, the control software may detect an error and perform error processing even though the x-ray output is normal.
What should I do if I have this product?
Voluntary: Firm initiated · Letter