VOCSN Patient Breathing Package (Pediatric, Active, Oxygen, Blue) REF PRT-01013-000 with pouch REF PRT-00802-001). Used with the VOCSN Unified Respiratory System which provides continuous or intermittent ventilatory support for the care of individuals who require mechanical ventilation.
Published March 6, 2024 · Recall initiated: January 9, 2024
Summary
FDA records medical device Class I recall Z-1070-2024, published March 6, 2024 and initiated January 9, 2024. Recalling firm: Ventec Life Systems, Inc.; Status: Ongoing; Reported quantity: 250 breathing circuits; Lot: Box Label - Part Numbers: PRT-01013-000 Pouch Label - Part Number: PRT-00802-001 UDI-DI Code: 00855573007594 Lot Number: 220922 Pack Slip Lot Number: 257337. Class I means a reasonable probability of serious adverse health consequences or death.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The bonded spiral wrap may detach before or during ventilation due to a manufacturing issue. Detachment of the spiral wrap can compromise structural integrity, functionality and/ or performance resulting in occlusions or leaks of the breathing circuit which may cause serious injury or death. The risk of injury or death is increased for ventilator dependent patients.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Ventec Life Systems, Inc.
- Distributor
- U.S.: CA, CO, FL, HI, IN, MO, NC, and WA O.U.S.: Japan
- Lot
- Box Label - Part Numbers: PRT-01013-000 Pouch Label - Part Number: PRT-00802-001 UDI-DI Code: 00855573007594 Lot Number: 220922 Pack Slip Lot Number: 257337
- Risk classification
- Class I
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAVC(Japan) Model Number: PRT-01098-100, Package PN:PRT-01125-100. used for ventilatory support.October 28, 2026
- FDAVC+Pro (Japan) Model Number: PRT-00900-100, Package PN:PRT-00973-100. used for ventilatory support.October 28, 2026
- FDAVCSN+Pro Model Number: PRT-01184-000, Package PN:PRT-01172-000. used for ventilatory support.October 28, 2026
- FDAVOCSN+Pro (Japan), Model Number: PRT-00490-100, Package PN:PRT-00853-100. used for ventilatory support.October 28, 2026
- FDAAirLife Adult Manual Resuscitator, 40" (1.0m) Oxygen Reservoir Tubing, Adult Mask, REF 2K8004March 6, 2024
- FDAMEDLINE TRAY TRACH BASIC SOLUTION, REF DYND40511March 6, 2024
Questions about this recall
Was VOCSN Patient Breathing Package (Pediatric, Active, Oxygen, Blue) REF PRT-01013-000 with pouch REF PRT-00802-001). Used with the VOCSN Unified Respiratory System which provides continuous or intermittent ventilatory support for the care of individuals who require mechanical ventilation. recalled?
Yes. FDA published recall Z-1070-2024 on March 6, 2024. Reported hazard: The bonded spiral wrap may detach before or during ventilation due to a manufacturing issue. Detachment of the spiral wrap can compromise structural integrity, functionality and/ or performance resulting in occlusions or leaks of the breathing circuit which may cause serious injury or death. The risk of injury or death is increased for ventilator dependent patients.
What should I do if I have this product?
Voluntary: Firm initiated · Letter