Microbiologics Enteric Pathogens Control Panel (Inactivated Pellet), REF 8184.
Published February 21, 2024 · Recall initiated: January 8, 2024
Summary
FDA records medical device Class II recall Z-1075-2024, published February 21, 2024 and initiated January 8, 2024. Recalling firm: Microbiologics Inc; Status: Ongoing; Reported quantity: 580 packages; Lot: Lot #8184-68, Exp. 5/31/2025, UDI-DI 70845357041448. UPDATE: The recall was expanded on 1/23/2024 to include the following additional lot numbers: 8184-65, exp. 3/31/2025; 8184-66, exp. 4/30/2025; 8184-67, exp. 5/31/2025; 8184-69, exp. 6/30/2025; 8184-70, exp. 8/31/2025; and 8184-71, exp. 8/31/2025. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The negative control was contaminated with one of the positive control pathogens (Norovirus). This would result in users getting a positive call for Norovirus when running the negative control.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Microbiologics Inc
- Distributor
- US Nationwide distribution in the states of AL, AZ, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, MA, MI, MN, MO, MS, MT, NE, NH, NJ, NY, OH, PA, RI, TN, TX, VA, WA, WI, WV, and WY. There was government distribution and no military distribution. Foreign distribution was made to France.
- Lot
- Lot #8184-68, Exp. 5/31/2025, UDI-DI 70845357041448. UPDATE: The recall was expanded on 1/23/2024 to include the following additional lot numbers: 8184-65, exp. 3/31/2025; 8184-66, exp. 4/30/2025; 8184-67, exp. 5/31/2025; 8184-69, exp. 6/30/2025; 8184-70, exp. 8/31/2025; and 8184-71, exp. 8/31/2025.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAKWIK-STIK, Quality control kit for culture media, Vibrio parahaemolyticus derived from ATCC 17802, CatalogDecember 10, 2025
- FDALYFO-DISK, M. Canis Derived from ATCC 36299, packaged as: 1) REF 0894K (6-pack), 2) REF 0894P (2-pack)August 6, 2025
- FDAKWIK-STICK, Quality control kit for culture media, Catalog Numbers 01065K, Oligella urethralis derived fromJune 18, 2025
- FDAMicrobiologics Multi-Strain Human Papillomavirus (HPV) Verification Panel (Inactivated Pellet), REF 8235;February 19, 2025
- FDAImpella catheters - Intravascular micro axial blood pumps that support a patient's circulatory system. (1)February 21, 2024
- FDAModel No. BEQ-HLS 5050 USA, HLS Set Advanced 5.0, Product Code/Part Number 70106.9077. Used with theFebruary 21, 2024
Questions about this recall
Was Microbiologics Enteric Pathogens Control Panel (Inactivated Pellet), REF 8184. recalled?
Yes. FDA published recall Z-1075-2024 on February 21, 2024. Reported hazard: The negative control was contaminated with one of the positive control pathogens (Norovirus). This would result in users getting a positive call for Norovirus when running the negative control.
What should I do if I have this product?
Voluntary: Firm initiated · Letter