Cobas b 123 POC system The cobas b 123 POC system is a fully automated POC system for whole blood in vitro measurement of pH, blood gases (B3G), electrolytes Nat, K+, iCaWt (ISE), hematocrit (THct), metabolites (Glu, Lac), total hemoglobin (tHb), hemoglobin derivatives (O2Hb, HHb, COT~b, MetHb), and oxygen saturation (SO2). In addition, the cobas b 123 POC system calculates derived parameters. It is dedicated for use in a Point-of-Care environment and laboratory. The integrated AutoQC module and the oximeter module are available as an option

Published February 1, 2017 · Recall initiated: September 3, 2015

Summary

FDA records medical device Class II recall Z-1077-2017, published February 1, 2017 and initiated September 3, 2015. Recalling firm: Roche Diagnostics Operations, Inc.; Status: Terminated; Reported quantity: 30; Lot: Models: cobas b 123 <2> POC system: 05122252001 cobas b 123 < 4 > POC system: 05122287001 Serial numbers: 11074 2274 2550 2302 2304 2309 1253 2420 2337 2147 2389 2418 2425 1749 2462 2338 2339 1850 2328 2500 1374 2301 1349 2400 2344 2350 1860 1941 2006 2128. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

under specific settings, an issue may occur during simultaneous Sensor Cartridge and Fluid Pack change on the cobas b 123 <2> POC system and cobas b 123 <4> POC system. The issue occurs when the software function [AutoQC as follow-up] is configured to run all three levels of AutoQC only after a Fluid Pack change, but not after a Sensor Cartridge change. When both are changed simultaneously, starting with the Sensor Cartridge and followed by the Fluid Pack, the analyzer carries out only the follow-up actions associated with the Sensor Cartridge change after completing the change workflow. As a result, no follow-up AutoQC is performed and the three expected AutoQC measurements for the Fluid Pack change are not carried out. Without running quality control, there is a remote possibility that system issues would not be detected and wrong results would not be excluded on all parameters: pH, PO2, PCO2, Na+, K+, Ca++, Cl-, Glu, Lac, Hct, SO2, O2Hb, COHb, MetHb, HHb, and Bili.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Roche Diagnostics Operations, Inc.
Distributor
Nationwide Distribution including CA, LA, NE, PR, SC, TX, VA, and WV
Lot
Models: cobas b 123 <2> POC system: 05122252001 cobas b 123 < 4 > POC system: 05122287001 Serial numbers: 11074 2274 2550 2302 2304 2309 1253 2420 2337 2147 2389 2418 2425 1749 2462 2338 2339 1850 2328 2500 1374 2301 1349 2400 2344 2350 1860 1941 2006 2128
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Cobas b 123 POC system The cobas b 123 POC system is a fully automated POC system for whole blood in vitro measurement of pH, blood gases (B3G), electrolytes Nat, K+, iCaWt (ISE), hematocrit (THct), metabolites (Glu, Lac), total hemoglobin (tHb), hemoglobin derivatives (O2Hb, HHb, COT~b, MetHb), and oxygen saturation (SO2). In addition, the cobas b 123 POC system calculates derived parameters. It is dedicated for use in a Point-of-Care environment and laboratory. The integrated AutoQC module and the oximeter module are available as an option recalled?

Yes. FDA published recall Z-1077-2017 on February 1, 2017. Reported hazard: under specific settings, an issue may occur during simultaneous Sensor Cartridge and Fluid Pack change on the cobas b 123 <2> POC system and cobas b 123 <4> POC system. The issue occurs when the software function [AutoQC as follow-up] is configured to run all three levels of AutoQC only after a Fluid Pack change, but not after a Sensor Cartridge change. When both are changed simultaneously, starting with the Sensor Cartridge and followed by the Fluid Pack, the analyzer carries out only the follow-up actions associated with the Sensor Cartridge change after completing the change workflow. As a result, no follow-up AutoQC is performed and the three expected AutoQC measurements for the Fluid Pack change are not carried out. Without running quality control, there is a remote possibility that system issues would not be detected and wrong results would not be excluded on all parameters: pH, PO2, PCO2, Na+, K+, Ca++, Cl-, Glu, Lac, Hct, SO2, O2Hb, COHb, MetHb, HHb, and Bili.

What should I do if I have this product?

Voluntary: Firm initiated · Letter