FDAFDA · Food and drugs·Recall number Z-1079-2014

Trabecular Metal" Reverse Shoulder System

Published March 5, 2014 · Recall initiated: October 17, 2013

Summary

FDA records medical device Class II recall Z-1079-2014, published March 5, 2014 and initiated October 17, 2013. Recalling firm: Zimmer, Inc.; Status: Terminated; Reported quantity: 73,002 units; Lot: 61025613, 61009830, 61038635, 61031385, 61021507, 61021506, 61016907, 61009834, 61006493, 61006492, 60999239, 60993837, 60993836, 60987295, 60982839, 60982838, 60970974, 60970973, 60950253, 60964425, 60946627, 60970946, 60964426, 60943695, 60940448, 60950251, 60946624, 60940449, 60927405, 60934492, 60900542, 60927406, 60919029, 60914617, 60919028, 60900543, 60893180, 60893181, 60906211, 60887745, 60887744, 60875337, 60875336, 60868040, 60873954, 60868041, 60867386, 60867385, 60856882, 60856880, 60855532, 60852815, 60855529, 60846158, 60846157, 60840619, 60834839, 60806512, 60824952, 60820211, 60814537, 60785523, 60795705, 60792083, 60761203, 60752933, 60781096, 60776583, 60765951, 60768804, 60757206, 60747093, 60741590, 60747092, 60741589, 60738183, 60727960, 60738184, 60730072, 60730073, 60723302, 60723277, 60718989, 60712568, 60683321, 60675599, 60669310, 60674468, 105533, 60605155, 105588, 60577323, 60580942, 60572796, 60564704, 60572798, 60568536, 60568535, 60578677, 60568468, 60568467, 60568466, 60564623, 60564639, 60564628, 60564641, 60564638, 60564636, 60568470, 60564642, 60564635, 60564625, 60564621, and 60564633. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

As a result of complaints for disassociation of the TM Reverse Glenosphere from the Baseplate, Zimmer updated the related surgical technique and released a video to provide additional instructions to users on how to properly assemble the Glenosphere to the Baseplate.

What to do

Voluntary: Firm initiated · Other

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Zimmer, Inc.
Distributor
Worldwide Distribution-USA (nationwide) and the countries of Germany, India, Malaysia, China, Singapore, Korea, Taiwan, and Brazil.
Lot
61025613, 61009830, 61038635, 61031385, 61021507, 61021506, 61016907, 61009834, 61006493, 61006492, 60999239, 60993837, 60993836, 60987295, 60982839, 60982838, 60970974, 60970973, 60950253, 60964425, 60946627, 60970946, 60964426, 60943695, 60940448, 60950251, 60946624, 60940449, 60927405, 60934492, 60900542, 60927406, 60919029, 60914617, 60919028, 60900543, 60893180, 60893181, 60906211, 60887745, 60887744, 60875337, 60875336, 60868040, 60873954, 60868041, 60867386, 60867385, 60856882, 60856880, 60855532, 60852815, 60855529, 60846158, 60846157, 60840619, 60834839, 60806512, 60824952, 60820211, 60814537, 60785523, 60795705, 60792083, 60761203, 60752933, 60781096, 60776583, 60765951, 60768804, 60757206, 60747093, 60741590, 60747092, 60741589, 60738183, 60727960, 60738184, 60730072, 60730073, 60723302, 60723277, 60718989, 60712568, 60683321, 60675599, 60669310, 60674468, 105533, 60605155, 105588, 60577323, 60580942, 60572796, 60564704, 60572798, 60568536, 60568535, 60578677, 60568468, 60568467, 60568466, 60564623, 60564639, 60564628, 60564641, 60564638, 60564636, 60568470, 60564642, 60564635, 60564625, 60564621, and 60564633.
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Trabecular Metal" Reverse Shoulder System recalled?

Yes. FDA published recall Z-1079-2014 on March 5, 2014. Reported hazard: As a result of complaints for disassociation of the TM Reverse Glenosphere from the Baseplate, Zimmer updated the related surgical technique and released a video to provide additional instructions to users on how to properly assemble the Glenosphere to the Baseplate.