Medtronic MiniMed Paradigm Insulin Infusion Pumps Model Numbers: MMT-511, MMT-512, MMT-712, MMT-712E, MMT-515, MMT-715, MMT-522, MMT-522K, MMT-722, MMT-722K, MMT-523, MMT-523K, MMT-723, MMT-723K, MMT-554, and MMT-754. Product Usage: Paradigm insulin infusion pumps are intended for use to deliver basal and bolus insulin to persons with diabetes.
Published April 17, 2013 · Recall initiated: March 22, 2013
Summary
FDA records medical device Class II recall Z-1085-2013, published April 17, 2013 and initiated March 22, 2013. Recalling firm: Medtronic MiniMed; Status: Terminated; Reported quantity: 428,000 units; Lot: Not available. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Medtronic is recalling the Medtronic Paradigm Insulin Pump because the pump's drive support cap may become detached from the pump case and protrude from the lower right side of the pump. When the pump is expose to water it may result in damage to the pump's internal electronics. This moisture damage can prevent the pump's buttons from working properly or can cause the pump to alarm.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Medtronic MiniMed
- Distributor
- Worldwide Distribution - USA (nationwide) the countries of: AE, AR, AT, AU, BA, BH, BM, BR, BS, CA, CH, CL, CN, CO, CR, CY, CZ, DE, DK, DO, DZ, EC, EE, EG, ES, FI, FR, GB, GR, GT, HK, HR, HU, IE, IL, IN, IQ, IT, JO, JP, KR, KW, LB, LC, LV, MK, MX, MY, NL, NO, NZ, OM, PA, PE, PL, PR, PY, QA, RO, SA, SE, SG, SI, SK, SV, SY, TF, TH, TR, TT, TW, UA, UY, VE, ZA.
- Lot
- Not available
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAMedtronic MiniMed Paradigm Insulin pump, model number MMT-523KApril 1, 2020
- FDAMiniMed Insulin Pump, model # MMT-508April 1, 2020
- FDAMedtronic MiniMed Paradigm Insulin pump, model number MMT-522April 1, 2020
- FDAMedtronic MiniMed Paradigm Insulin pump, model number MMT-723April 1, 2020
- FDABracket Buccal Tube U7L DB/UL22 Roth MT HK 0/-14/0 Product Usage: Bracket Buccal Tubes are intended to hold inApril 17, 2013
- FDALHB, Type I First Aid Kit Perishable, Non-Aspirin 4/pg, Aspirin 4/pg and Povidone Iodine, antisepticApril 17, 2013
Questions about this recall
Was Medtronic MiniMed Paradigm Insulin Infusion Pumps Model Numbers: MMT-511, MMT-512, MMT-712, MMT-712E, MMT-515, MMT-715, MMT-522, MMT-522K, MMT-722, MMT-722K, MMT-523, MMT-523K, MMT-723, MMT-723K, MMT-554, and MMT-754. Product Usage: Paradigm insulin infusion pumps are intended for use to deliver basal and bolus insulin to persons with diabetes. recalled?
Yes. FDA published recall Z-1085-2013 on April 17, 2013. Reported hazard: Medtronic is recalling the Medtronic Paradigm Insulin Pump because the pump's drive support cap may become detached from the pump case and protrude from the lower right side of the pump. When the pump is expose to water it may result in damage to the pump's internal electronics. This moisture damage can prevent the pump's buttons from working properly or can cause the pump to alarm.
What should I do if I have this product?
Voluntary: Firm initiated · Letter