The Rapid fFN Specimen Collection Tube (PN: 71550-001) is a component of the Rapid fFN Specimen Collection kit (PN: 71738-001). The Collection Tube (PN: 71550-001) is manufactured by a third-party supplier, MML Diagnostics Packaging, and includes the transport tube, cap, and extraction buffer. Tubes are packaged into PN: 71738-00 at Hologic prior to kit release.
Published February 12, 2025 · Recall initiated: January 8, 2025
Summary
FDA records medical device Class II recall Z-1095-2025, published February 12, 2025 and initiated January 8, 2025. Recalling firm: Hologic, Inc.; Status: Ongoing; Reported quantity: 1351 units; Lot: Lot # 901164, 901165, 904850/ UDI: 25420045514765. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Three lots of Rapid fFN Specimen Collection kit (901164, 901165, and 904850) contain Rapid fFN Specimen Collection Tubes which may be empty, underfilled, or overfilled with extraction buffer. The transport tubes are used with the Fetal Fibronectin Enzyme Immunoassay and Rapid fFN® for the TLiIQ® System which aids in assessing the risk of preterm delivery. The use of empty transport tubes or transport tubes containing insufficient extraction buffer volume may result in invalid results and/or delays in obtaining results. Additionally, the use of underfilled tubes may result in false positive results and the use of overfilled tubes may result in false negative results.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Hologic, Inc.
- Distributor
- Worldwide - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, Guam and the countries of Canada and Japan.
- Lot
- Lot # 901164, 901165, 904850/ UDI: 25420045514765
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAAzurion 3 M12 System Code: (1) 722063 (2) 722221February 12, 2025
- FDAIntelliVue Patient Monitor MX550. Model Number: 866066February 12, 2025
- FDAPanther Fusion SARS-CoV-2/Flu A/B/RSV Assay, REF: PRD-07400January 1, 2025
- FDAAptima CMV Quant Assay, Reference numbers: AW-22600-001, AW-25509-001, and AW-26381-001 Amplification Reagent,April 3, 2024
- FDAPANTHER FUSION System REF 9121010000 The Panther System is an integrated nucleic acid testing system whichJanuary 31, 2024
- FDASelenia Dimensions Mammography System, Model number ASY-04160, Selenia Dimensions Gantry, UDI 15420045505636,March 4, 2020
Questions about this recall
Was The Rapid fFN Specimen Collection Tube (PN: 71550-001) is a component of the Rapid fFN Specimen Collection kit (PN: 71738-001). The Collection Tube (PN: 71550-001) is manufactured by a third-party supplier, MML Diagnostics Packaging, and includes the transport tube, cap, and extraction buffer. Tubes are packaged into PN: 71738-00 at Hologic prior to kit release. recalled?
Yes. FDA published recall Z-1095-2025 on February 12, 2025. Reported hazard: Three lots of Rapid fFN Specimen Collection kit (901164, 901165, and 904850) contain Rapid fFN Specimen Collection Tubes which may be empty, underfilled, or overfilled with extraction buffer. The transport tubes are used with the Fetal Fibronectin Enzyme Immunoassay and Rapid fFN® for the TLiIQ® System which aids in assessing the risk of preterm delivery. The use of empty transport tubes or transport tubes containing insufficient extraction buffer volume may result in invalid results and/or delays in obtaining results. Additionally, the use of underfilled tubes may result in false positive results and the use of overfilled tubes may result in false negative results.
What should I do if I have this product?
Voluntary: Firm initiated · Letter