Back to search
FDAFDA · Food and drugs·Recall number Z-1095-2025

The Rapid fFN Specimen Collection Tube (PN: 71550-001) is a component of the Rapid fFN Specimen Collection kit (PN: 71738-001). The Collection Tube (PN: 71550-001) is manufactured by a third-party supplier, MML Diagnostics Packaging, and includes the transport tube, cap, and extraction buffer. Tubes are packaged into PN: 71738-00 at Hologic prior to kit release.

Published February 12, 2025 · Recall initiated: January 8, 2025

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Three lots of Rapid fFN Specimen Collection kit (901164, 901165, and 904850) contain Rapid fFN Specimen Collection Tubes which may be empty, underfilled, or overfilled with extraction buffer. The transport tubes are used with the Fetal Fibronectin Enzyme Immunoassay and Rapid fFN¿ for the TLiIQ¿ System which aids in assessing the risk of preterm delivery. The use of empty transport tubes or transport tubes containing insufficient extraction buffer volume may result in invalid results and/or delays in obtaining results. Additionally, the use of underfilled tubes may result in false positive results and the use of overfilled tubes may result in false negative results.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Hologic, Inc.
Distributor
Worldwide - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, Guam and the countries of Canada and Japan.
Lot
Lot # 901164, 901165, 904850/ UDI: 25420045514765
Risk classification
Class II
Area (FDA)
device
Country
United States

Sell products like this?

Recalert watches your whole catalog for recalls like this one, automatically.

Start free — no card