MRT-1504/A4 - Titan X MRI System MRT-30 I0A/5 - Titan 3T MRI System Product Usage: MRI System.
Published April 17, 2013 · Recall initiated: October 15, 2012
Summary
FDA records medical device Class II recall Z-1101-2013, published April 17, 2013 and initiated October 15, 2012. Recalling firm: Toshiba American Medical Systems Inc; Status: Terminated; Reported quantity: 27 units; Lot: Product Numbers: MRT-1504/A4, MRT-1504/S4, MRT-3010/A5, MZKT-UPKHE/S1. Serial Numbers unique to each consignee. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
It was found that there is a problem with the LCD monitor used in the console. The LCD monitor may become blank during use. Specifically, the backlight of the LCD monitor may turn off while the system is in use, causing the LCD monitor to become blank. However, the other parts of the system continue to operate normally, and the risk of a serious health hazard is therefore very low. The LCD Moni
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Toshiba American Medical Systems Inc
- Distributor
- USA Nationwide Distribution including the states of : IL, NH, MA, MI, MO, PR, NV, NM, MN, ID, NY, OH, NC, AZ, MS, LA, OR, PA, CA, and CT.
- Lot
- Product Numbers: MRT-1504/A4, MRT-1504/S4, MRT-3010/A5, MZKT-UPKHE/S1. Serial Numbers unique to each consignee.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAINFX-8000C Fluoroscopic X-Ray SystemsMarch 14, 2018
- FDAAPLIO 500 TUS-A500 diagnostic ultrasound systemJuly 19, 2017
- FDAXARIO 200 TUS-X200 diagnostic ultrasound systemJuly 19, 2017
- FDAToshiba Kalare DREX-KL80 X-ray generatorMarch 22, 2017
- FDAbuPROPion Hydrochloride Extended-Release Tablets (XL) 300 mg, 30-count tablets per bottle, Rx only,April 24, 2013
- FDAGE Healthcare, Discovery MR750w 3.0T System. Part Number: 5352293. Product Usage: It is indicated for use as aApril 17, 2013
Questions about this recall
Was MRT-1504/A4 - Titan X MRI System MRT-30 I0A/5 - Titan 3T MRI System Product Usage: MRI System. recalled?
Yes. FDA published recall Z-1101-2013 on April 17, 2013. Reported hazard: It was found that there is a problem with the LCD monitor used in the console. The LCD monitor may become blank during use. Specifically, the backlight of the LCD monitor may turn off while the system is in use, causing the LCD monitor to become blank. However, the other parts of the system continue to operate normally, and the risk of a serious health hazard is therefore very low. The LCD Moni
What should I do if I have this product?
Voluntary: Firm initiated · Letter