MEDLINE Kits, trays, and packs labeled as follows: a) ASC MINOR THYROID CDS, REF CDS983682F; b) BRONCHOSCOPY KIT, REF DYKE1718A; c) BRONCHOSCOPY KIT, REF DYKE1718B; d) DOUBLE CENTRAL LINE CAP CHANGE, REF DYNDC2214G; e) ENT PACK, REF DYNJ55334C; f) ENT PACK, REF DYNJ45175B; g) ENT PACK, REF DYNJ59030A; h) ENT PACK, REF DYNJ83416; i) ENT/EPISTAXIS TRAY, REF MNS2895; j) ENT/EPISTAXIS TRAY, REF MNS2895H; k) HEAD & NECK PACK, REF DYNJ60166C; l) HEAD AND NECK CDS, REF CDS983910K; m) HEAD AND NECK PACK, REF DYNJ68872C; n) HEAD AND NECK PACK, REF DYNJ46216J; o) NASAL, REF CDS984899G; p) NASAL, REF CDS984899I; q) PACK,BRONCHOSCOPY, REF DYNJ65475A; r) SINGLE CENTRAL LINE CAP CHANGE, REF DYNDC2213F; s) T & A, REF CDS984895I; t) T AND A, REF DYNJ36719C; u) T AND A, REF DYNJ906741B; v) T AND A CDS, REF CDS983362C; w) T&A PACK, REF DYNJ33112I; x) TONSIL CDS-LF, REF CDS980272L; y) TRACH CARE KIT, REF DYNDJN001A
Published March 6, 2024 · Recall initiated: November 15, 2023
Summary
FDA records medical device Class I recall Z-1101-2024, published March 6, 2024 and initiated November 15, 2023. Recalling firm: MEDLINE INDUSTRIES, LP - Northfield; Status: Ongoing; Lot: a) REF CDS983682F; UDI/DI 40193489852975 (case), 10193489852974 (unit), lot Numbers: 22DDA235, 22GBJ113, 23ABQ437, 23EBS939; b) REF DYKE1718A; UDI/DI 40193489494656 (case), 10193489494655 (unit), Lot Numbers: 22FBU098, 22HBM601, 23ABM389; c) REF DYKE1718B; UDI/DI 40195327240128 (case), 10195327240127 (unit), Lot Numbers: 23DBE977; d) REF DYNDC2214G; UDI/DI 40195327051984 (case), 10195327051983 (unit), Lot Numbers: 22ABP471, 22BBO794, 22OBJ991; e) REF DYNJ55334C; UDI/DI 40889942703211 (case), 10889942703210 (unit), Lot Numbers: 22ABL051, 22EBV190, 22FBS394, 23BBA777, 23GBL169; f) REF DYNJ45175B; UDI/DI 40889942485216 (case), 10889942485215 (unit), Lot Numbers: 22CLB148, 22GLB002; g) REF DYNJ59030A; UDI/DI 40195327026203 (cases), 10195327026202 (unit), Lot Numbers: 22DBR838, 22EBO071, 22JBW765, 23FBF261, 23GBC456, 23HBR618; h) REF DYNJ83416; UDI/DI 40195327265053 (cases), 10195327265052 (unit), Lot Numbers: 22LBF471, 23ABN102, 23BBA471; i) REF MNS2895; UDI/DI 50653160062934 (case), 653160062939 (unit), Lot Numbers: 23BBQ180, 23DBN766, 23EBH195, 23FBE540; j) REF MNS2895H; UDI/DI 50653160062934 (case), 653160062939 (unit), Lot Numbers: 23BBQ180, 23DBN766, 23EBH195, 23FBE540; k) REF DYNJ60166C; UDI/DI 40193489467148 (case), 10193489467147 (unit), Lot Numbers: 23ELB099; l) REF CDS983910K; UDI/DI 40195327256594 (case), 10195327256593 (unit), Lot Numbers: 22JBT244, 23ABP271, 23ABP272, 23ABQ792, 23CBE912, 23DBR779, 23FBB371, 23JBL006, 23JBR337; m) REF DYNJ68872C; UDI/DI 40195327375714 (case), 10195327375713 (unit), Lot Numbers: 23DBC646, 23FBA039, 23GBA604, 23HBZ389, 23JBN556; n) REF DYNJ46216J; UDI/DI 40195327407491 (case), 10195327407490 (unit), Lot Numbers: 23EBF873, 23GBO919, 23GBT576, 23IBE145; o) REF CDS984899G; UDI/DI 40193489983822 (case), 10193489983821 (unit), Lot Numbers: 22CLA789; p) REF CDS984899I; UDI/DI 40195327313297 (case), 10195327313296 (unit), Lot Numbers: 23IBV180; q) REF DYNJ65475A; UDI/DI 40193489974578 (case), 10193489974577 (unit), Lot Numbers: 22DBF905; r) REF DYNDC2213F; UDI/DI 40195327052080 (case), 10195327052089 (unit), Lot number: 22ABM936, 22CMG684, 22EMB991; s) REF CDS984895I; UDI/DI 40193489983815 (case), 10193489983814 (unit), Lot Number: 22FBL382, 22IBD790, 23ABM451, 23DBM334; t) REF DYNJ36719C; UDI/DI 40193489265539 (case), 10193489265538 (unit), Lot Numbers: 21LBA818, 21LBU235, 22BBL912; u) REF DYNJ906741B; UDI/DI 40195327055364 (case), 10195327055363 (unit), Lot Numbers: 22ABF704, 22ABL214, 22BBQ966, 22FBY564, 22GBQ902, 22HMI302, 23AMA387, 23AMI317; v) REF CDS983362C; UDI/DI 40193489267342 (case), 10193489267341 (unit), Lot Numbers: 23ABI733; w) REF DYNJ33112I; UDI/DI 40193489402736 (case), 10193489402735 (unit), Lot Numbers: 22DBO140, 22DBO155, 22JBY668, 23BBT134, 23DBT051, 23EBA549, 23HBI328, 23IBM005; x) REF CDS980272L; UDI/DI 40193489616331 (case), 10193489616330 (unit), Lot Numbers: 22NBC732; y) REF DYNDJN001A; UDI/DI 40193489421614 (case), 10193489421613 (unit), Lot Numbers: 22EBB217, 22FBQ070, 22FBU496. Class I means a reasonable probability of serious adverse health consequences or death.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Medline Industries, LP is recalling certain kits and trays that were manufactured using specific lots of Nurse Assist (as components): 0.9% Sodium Chloride Irrigation USP, Sterile Water for Irrigation USP and Saline Flush Syringe. Nurse Assist issued a recall due to the potential lack of sterility, which could result in the solution being non sterile.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- MEDLINE INDUSTRIES, LP - Northfield
- Distributor
- Distribution US nationwide including Puerto Rico, Canada, UAE, and Panama
- Lot
- a) REF CDS983682F; UDI/DI 40193489852975 (case), 10193489852974 (unit), lot Numbers: 22DDA235, 22GBJ113, 23ABQ437, 23EBS939; b) REF DYKE1718A; UDI/DI 40193489494656 (case), 10193489494655 (unit), Lot Numbers: 22FBU098, 22HBM601, 23ABM389; c) REF DYKE1718B; UDI/DI 40195327240128 (case), 10195327240127 (unit), Lot Numbers: 23DBE977; d) REF DYNDC2214G; UDI/DI 40195327051984 (case), 10195327051983 (unit), Lot Numbers: 22ABP471, 22BBO794, 22OBJ991; e) REF DYNJ55334C; UDI/DI 40889942703211 (case), 10889942703210 (unit), Lot Numbers: 22ABL051, 22EBV190, 22FBS394, 23BBA777, 23GBL169; f) REF DYNJ45175B; UDI/DI 40889942485216 (case), 10889942485215 (unit), Lot Numbers: 22CLB148, 22GLB002; g) REF DYNJ59030A; UDI/DI 40195327026203 (cases), 10195327026202 (unit), Lot Numbers: 22DBR838, 22EBO071, 22JBW765, 23FBF261, 23GBC456, 23HBR618; h) REF DYNJ83416; UDI/DI 40195327265053 (cases), 10195327265052 (unit), Lot Numbers: 22LBF471, 23ABN102, 23BBA471; i) REF MNS2895; UDI/DI 50653160062934 (case), 653160062939 (unit), Lot Numbers: 23BBQ180, 23DBN766, 23EBH195, 23FBE540; j) REF MNS2895H; UDI/DI 50653160062934 (case), 653160062939 (unit), Lot Numbers: 23BBQ180, 23DBN766, 23EBH195, 23FBE540; k) REF DYNJ60166C; UDI/DI 40193489467148 (case), 10193489467147 (unit), Lot Numbers: 23ELB099; l) REF CDS983910K; UDI/DI 40195327256594 (case), 10195327256593 (unit), Lot Numbers: 22JBT244, 23ABP271, 23ABP272, 23ABQ792, 23CBE912, 23DBR779, 23FBB371, 23JBL006, 23JBR337; m) REF DYNJ68872C; UDI/DI 40195327375714 (case), 10195327375713 (unit), Lot Numbers: 23DBC646, 23FBA039, 23GBA604, 23HBZ389, 23JBN556; n) REF DYNJ46216J; UDI/DI 40195327407491 (case), 10195327407490 (unit), Lot Numbers: 23EBF873, 23GBO919, 23GBT576, 23IBE145; o) REF CDS984899G; UDI/DI 40193489983822 (case), 10193489983821 (unit), Lot Numbers: 22CLA789; p) REF CDS984899I; UDI/DI 40195327313297 (case), 10195327313296 (unit), Lot Numbers: 23IBV180; q) REF DYNJ65475A; UDI/DI 40193489974578 (case), 10193489974577 (unit), Lot Numbers: 22DBF905; r) REF DYNDC2213F; UDI/DI 40195327052080 (case), 10195327052089 (unit), Lot number: 22ABM936, 22CMG684, 22EMB991; s) REF CDS984895I; UDI/DI 40193489983815 (case), 10193489983814 (unit), Lot Number: 22FBL382, 22IBD790, 23ABM451, 23DBM334; t) REF DYNJ36719C; UDI/DI 40193489265539 (case), 10193489265538 (unit), Lot Numbers: 21LBA818, 21LBU235, 22BBL912; u) REF DYNJ906741B; UDI/DI 40195327055364 (case), 10195327055363 (unit), Lot Numbers: 22ABF704, 22ABL214, 22BBQ966, 22FBY564, 22GBQ902, 22HMI302, 23AMA387, 23AMI317; v) REF CDS983362C; UDI/DI 40193489267342 (case), 10193489267341 (unit), Lot Numbers: 23ABI733; w) REF DYNJ33112I; UDI/DI 40193489402736 (case), 10193489402735 (unit), Lot Numbers: 22DBO140, 22DBO155, 22JBY668, 23BBT134, 23DBT051, 23EBA549, 23HBI328, 23IBM005; x) REF CDS980272L; UDI/DI 40193489616331 (case), 10193489616330 (unit), Lot Numbers: 22NBC732; y) REF DYNDJN001A; UDI/DI 40193489421614 (case), 10193489421613 (unit), Lot Numbers: 22EBB217, 22FBQ070, 22FBU496
- Risk classification
- Class I
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDACURAD Alcohol Prep Pads, Sterile, Medium, 2-Ply, Contents: 5 boxes per Carton, 30 Boxes per Case, Single UseSeptember 3, 2025
- FDAReliOn, Sterile Alcohol Swabs, Skin Cleanser, 200 Swabs, Distributed by, Walmart, Inc., Bentonville, AR 72716September 3, 2025
- FDAGood Neighbor Pharmacy, Alcohol Prep Pads, Distributed By AmeriSource Bergen, 1 West First Avenue,September 3, 2025
- FDAMedline Kits containing BD SureStep Foley Tray Systems: SKU/Description: CDS984076J GYN ROBOTIC DYNJG001004ASeptember 3, 2025
- FDAMEDLINE Kits, trays, and packs labeled as follows: SERUM TEAR KIT, REF DYNDA1545March 6, 2024
- FDAMEDLINE Kits, trays, and packs labeled as follows: a) ASC GENERAL MINOR CDS, REF CDS983673G; b) ASC GENERALMarch 6, 2024
Questions about this recall
Was MEDLINE Kits, trays, and packs labeled as follows: a) ASC MINOR THYROID CDS, REF CDS983682F; b) BRONCHOSCOPY KIT, REF DYKE1718A; c) BRONCHOSCOPY KIT, REF DYKE1718B; d) DOUBLE CENTRAL LINE CAP CHANGE, REF DYNDC2214G; e) ENT PACK, REF DYNJ55334C; f) ENT PACK, REF DYNJ45175B; g) ENT PACK, REF DYNJ59030A; h) ENT PACK, REF DYNJ83416; i) ENT/EPISTAXIS TRAY, REF MNS2895; j) ENT/EPISTAXIS TRAY, REF MNS2895H; k) HEAD & NECK PACK, REF DYNJ60166C; l) HEAD AND NECK CDS, REF CDS983910K; m) HEAD AND NECK PACK, REF DYNJ68872C; n) HEAD AND NECK PACK, REF DYNJ46216J; o) NASAL, REF CDS984899G; p) NASAL, REF CDS984899I; q) PACK,BRONCHOSCOPY, REF DYNJ65475A; r) SINGLE CENTRAL LINE CAP CHANGE, REF DYNDC2213F; s) T & A, REF CDS984895I; t) T AND A, REF DYNJ36719C; u) T AND A, REF DYNJ906741B; v) T AND A CDS, REF CDS983362C; w) T&A PACK, REF DYNJ33112I; x) TONSIL CDS-LF, REF CDS980272L; y) TRACH CARE KIT, REF DYNDJN001A recalled?
Yes. FDA published recall Z-1101-2024 on March 6, 2024. Reported hazard: Medline Industries, LP is recalling certain kits and trays that were manufactured using specific lots of Nurse Assist (as components): 0.9% Sodium Chloride Irrigation USP, Sterile Water for Irrigation USP and Saline Flush Syringe. Nurse Assist issued a recall due to the potential lack of sterility, which could result in the solution being non sterile.
What should I do if I have this product?
Voluntary: Firm initiated · Letter