Philips Azurion system; Software Version Number: R3.1;

Published January 28, 2026 · Recall initiated: December 12, 2025

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Philips has identified two (2) software issues affecting device systems that may result in loss of imaging (X-ray) functionality and/or loss of motorized movement, and/or incorrect image content and/or loss of data. Issue 1 - System remains in continuous restart mode after the start-up. Issue 2 - Longitudinal Position Error Applicable only to systems with Poly-G3 frontal stand.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Distributor
International distribution in the countries of AE, AR, AT, AZ, BE, BG, BH, BR, BS, CH, CL, CO, CR, CY, CZ, DE, DK, DZ, EC, EG, ES, GB, GR, GT, ID, IE, IL, IN, IQ, IT, JE, JP, LB, MA, MO, NI, NL, NO, PE, PL, RO, SA, SE, SV, TR, UA, VE, VN, ZA;
Lot
Software Version Number: R3.1; (1) Model Number: 722221; System Product Name: Azurion 3 M12; UDI-DI: 00884838099203; Serial Numbers: (2) Model Number: 722222; System Product Name: Azurion 3 M15; UDI-DI: 00884838099210; Serial Numbers: (3) Model Number: 722223; System Product Name: Azurion 7 M12; UDI-DI: 00884838099241; Serial Numbers: (4) Model Number: 722224; System Product Name: Azurion 7 M20; UDI-DI: 00884838099258; Serial Numbers: (5) Model Number: 722225; System Product Name: Azurion 7 B12; UDI-DI: 00884838099265; Serial Numbers: (6) Model Number: 722226; System Product Name: Azurion 7 B20; UDI-DI: 00884838099272; Serial Numbers: (7) Model Number: 722227; System Product Name: Azurion 5 M12; UDI-DI: 00884838099227; Serial Numbers: (8) Model Number: 722228; System Product Name: Azurion 5 M20; UDI-DI: 00884838099234; Serial Numbers: (9) Model Number: 722229; System Product Name: Azurion 3 M12; UDI-DI: 00884838116726; Serial Numbers: 5, 9, 32, 40, 35, 31, 33; (10) Model Number: 722230; System Product Name: Azurion 3 M15; UDI-DI: 00884838116733; Serial Numbers: 25, 54, 50, 52, 55, 30, 63, 60, 67; (11) Model Number: 722231; System Product Name: Azurion 5 M12; UDI-DI: 00884838116740; Serial Numbers: 62, 7, 8, 14, 17, 63, 6, 66, 59, 69, 56, 67, 48, 68, 80, 93, 87, 94, 81, 92, 18; (12) Model Number: 722232; System Product Name: Azurion 5 M20; UDI-DI: 00884838116757; Serial Numbers: 97, 39, 31, 100, 104, 93, 115, 113, 118, 119, 89, 54, 67, 7, 128, 86, 110, 132, 130, 131, 123, 137, 139, 117, 260; (13) Model Number: 722233; System Product Name: Azurion 7 M12; UDI-DI: 00884838116764; Serial Numbers: 239, 100, 233, 240, 176, 227, 223, 130, 161, 194, 243, 56, 193, 280, 75, 253, 156, 288, 72, 281, 304, 300, 105, 295, 312, 310, 285, 292, 290, 298, 297; (14) Model Number: 722234; System Product Name: Azurion 7 M20; UDI-DI: 00884838116771; Serial Numbers: 731, 727, 261, 473, 683, 722, 562, 750, 745, 766, 265, 701, 400, 310, 421, 314, 182, 579, 96, 763, 272, 612, 15, 844, 105, 702, 751, 384, 851, 869, 743, 337, 651, 455, 706, 312, 563, 709, 764, 871, 846, 278, 713, 802, 868, 277, 742, 184, 819, 759, 910, 928, 415, 414, 945, 961, 355, 933, 692, 920, 569, 939, 909, 553, 837, 856, 571, 756, 941, 570, 761, 981, 353, 888, 964, 820, 746, 736, 876, 599, 1017, 908, 487, 983, 495; (15) Model Number: 722235; System Product Name: Azurion 7 B12; UDI-DI: 00884838116788; Serial Numbers: 96, 79, 111, 100, 101, 38, 27, 92, 78, 118, 48, 131, 135, 22, 125, 103, 138, 139, 137; (16) Model Number: 722236; System Product Name: Azurion 7 B20; UDI-DI: 00884838116801; Serial Numbers: 241, 297, 198, 215, 305, 274, 299, 78, 279, 287, 318, 353, 324, 331, 293, 245, 314, 310, 317, 312, 77, 358, 206, 383, 380, 370, 377, 385, 405, 379, 328;
Risk classification
Class II
Area (FDA)
device
Country
Netherlands