FMP X-alt Acetabular Liner This acetabular liner is intended for use with DJO Surgical hip systems. It is used with an acetabular shell to replicate the kinematics of a natural hip.

Published March 12, 2014 · Recall initiated: February 7, 2014

Summary

FDA records medical device Class II recall Z-1130-2014, published March 12, 2014 and initiated February 7, 2014. Recalling firm: Djo Surgical; Status: Terminated; Reported quantity: 10 units of Model/Catalog Number 931-28-248 Lot/Serial Number(s) 685F1034 labeled asModel/Catalog Number 932-36-252, Lot# 728F1089; Lot: Model/Catalog Number 931-28-248 Lot/Serial Number 685F1034 (an MP5 Non-Hooded, Neutral Liner) were mistakenly packaged/labeled as Model/Catalog Number 932-36-252, Lot# 728F1089 (an MP7 10-degree Hooded Liner) and vice-versa. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Packaging error -- two different types and sizes of acetabular liners used during hip replacement surgery were mistakenly packaged into each other's outer boxes after sterilization, resulting in the products being mislabeled. The units labeled as Model/Catalog Number 931-28-248 Lot/Serial Number 685F1034 on the outside and containing Model/Catalog Number 932-36-252, Lot# 728F1089 on the inside (1

What to do

Voluntary: Firm initiated · Visit

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Djo Surgical
Distributor
Nationwide Distribution - USA including TX, LA, FL, CA, and AZ.
Lot
Model/Catalog Number 931-28-248 Lot/Serial Number 685F1034 (an MP5 Non-Hooded, Neutral Liner) were mistakenly packaged/labeled as Model/Catalog Number 932-36-252, Lot# 728F1089 (an MP7 10-degree Hooded Liner) and vice-versa
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was FMP X-alt Acetabular Liner This acetabular liner is intended for use with DJO Surgical hip systems. It is used with an acetabular shell to replicate the kinematics of a natural hip. recalled?

Yes. FDA published recall Z-1130-2014 on March 12, 2014. Reported hazard: Packaging error -- two different types and sizes of acetabular liners used during hip replacement surgery were mistakenly packaged into each other's outer boxes after sterilization, resulting in the products being mislabeled. The units labeled as Model/Catalog Number 931-28-248 Lot/Serial Number 685F1034 on the outside and containing Model/Catalog Number 932-36-252, Lot# 728F1089 on the inside (1

What should I do if I have this product?

Voluntary: Firm initiated · Visit