Back to search
FDAFDA · Food and drugs·Recall number Z-1133-2020

NC TREK Coronary Dilatation Catheter, REF 1012453-20 Rx Only, NC TREK RX 4.50 X 8MM BDC The NC TREK RX Coronary Dilatation Catheter is indicated for: a) balloon dilatation of the stenotic portion of a coronary artery or bypass graft stenosis, for the purpose of improving myocardial perfusion b) balloon dilatation of a coronary artery occlusion, for the purpose of restoring coronary flow in patients with ST-segment elevation myocardial infarction c) balloon dilatation of a stent after implantation (balloon models 2.00 mm 5.00 mm only)

Published February 26, 2020 · Recall initiated: January 29, 2020

Summary

FDA records medical device Class I recall Z-1133-2020, published February 26, 2020 and initiated January 29, 2020. Recalling firm: Abbott Vascular; Status: Terminated; Reported quantity: Total 39687 NC TREK units (13891 US and 25796 OUS); Lot: Device Identifier/GTIN: 08717648152092 Part Number: 1012454-08 Lot Numbers: 90801G1 90925G1 90812G1 91010G2 90818G1 91022G1 90818G2 91025G1 90904G1 91101G1 90912G1. Class I means a reasonable probability of serious adverse health consequences or death.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Coronary Dilatation Catheters may exhibit difficulty or inability to deflate the balloon.

What to do

Voluntary: Firm initiated · Telephone

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Abbott Vascular
Distributor
Worldwide Distribution: US (Nation-wide) and OUS (countries of): Algeria Jordan Argentina Korea Armenia Kuwait Aruba Lebanon Australia Malaysia Austria, Mauritius Bahrain Mexico Bangladesh Morocco Belgium Netherlands Brazil New Zealand Brunei Norway Bulgaria Oman Canada Pakistan Chile Panama China Peru Colombia Poland Costa Rica Portugal Croatia R®union Cyprus Romania Czech Republic Russian Federation Denmark Saudi Arabia Dominican Republic Singapore Egypt Slovakia Estonia Slovenia Finland South Africa France Spain Germany Sri Lanka Greece State of Palestine Hong Kong Sweden Hungary Switzerland Iceland Taiwan, Province of China India Thailand Indonesia Tunisia Iran Turkey Iraq Ukraine Ireland United Arab Emirates Israel United Kingdom Italy Uruguay Japan Viet Nam
Lot
Device Identifier/GTIN: 08717648152092 Part Number: 1012454-08 Lot Numbers: 90801G1 90925G1 90812G1 91010G2 90818G1 91022G1 90818G2 91025G1 90904G1 91101G1 90912G1
Risk classification
Class I
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was NC TREK Coronary Dilatation Catheter, REF 1012453-20 Rx Only, NC TREK RX 4.50 X 8MM BDC The NC TREK RX Coronary Dilatation Catheter is indicated for: a) balloon dilatation of the stenotic portion of a coronary artery or bypass graft stenosis, for the purpose of improving myocardial perfusion b) balloon dilatation of a coronary artery occlusion, for the purpose of restoring coronary flow in patients with ST-segment elevation myocardial infarction c) balloon dilatation of a stent after implantation (balloon models 2.00 mm 5.00 mm only) recalled?

Yes. FDA published recall Z-1133-2020 on February 26, 2020. Reported hazard: Coronary Dilatation Catheters may exhibit difficulty or inability to deflate the balloon.

What should I do if I have this product?

Voluntary: Firm initiated · Telephone