Medfusion Syringe Pump 3500 Series The Medfusion® Syringe Infusion Pumps are indicated for the following uses: 1) Administration of fluids requiring precisely controlled infusion rates including blood or blood products, lipids, drugs, antibiotics, enteral solutions and other therapeutic fluids; 2) By the following delivery routes: arterial, epidural, intravenous, intrathecal, subcutaneous, and enteral; 3)By the following delivery modes: continuous, volume/time, mass, body weight, intermittent, and bolus; 4) In critical care, anesthesia, neonatal, and pediatric applications or other healthcare settings where use of the syringe infusion pump can be monitored or supervised by a clinician; 5) Inside the MRI room mounted outside the 150 Gauss line and with shielded magnets of field strength of 1.5 Tesla.

Published March 28, 2018 · Recall initiated: November 13, 2017

Summary

FDA records medical device Class II recall Z-1135-2018, published March 28, 2018 and initiated November 13, 2017. Recalling firm: Smiths Medical ASD Inc.; Status: Terminated; Reported quantity: 16,600 pumps total; Lot: Product Codes 3500G, 3500VX, 3500-414, 3500-402, 3500-0600-249, 3500-0600-00, 3500-0600-01, 3500-0600-50, 3500-0600-51, 3500-0600-82, 3500-306, 3500-415, 3500-500. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Certain Medfusion Syringe Pump Models, Series 3100, 3500, and 4000, may not recognize or may misidentify loaded medication syringes. The inability of a pump to recognize a syringe (i.e. the size of the syringe is unknown to the pump) results in an inability to complete pump programming. Misidentification of a syringe is where the pump misinterprets the syringe size.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Smiths Medical ASD Inc.
Distributor
Worldwide distribution- US nationwide Distribution and to the countries of : Australia, Barbados, Canada, Hong Kong, Indonesia, Jamaica, New Zealand, Philippines, Saudi Arabia, Singapore, South Africa, Trinidad and Tobago, United Arab Emirates, and United Kingdom.
Lot
Product Codes 3500G, 3500VX, 3500-414, 3500-402, 3500-0600-249, 3500-0600-00, 3500-0600-01, 3500-0600-50, 3500-0600-51, 3500-0600-82, 3500-306, 3500-415, 3500-500
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Medfusion Syringe Pump 3500 Series The Medfusion® Syringe Infusion Pumps are indicated for the following uses: 1) Administration of fluids requiring precisely controlled infusion rates including blood or blood products, lipids, drugs, antibiotics, enteral solutions and other therapeutic fluids; 2) By the following delivery routes: arterial, epidural, intravenous, intrathecal, subcutaneous, and enteral; 3)By the following delivery modes: continuous, volume/time, mass, body weight, intermittent, and bolus; 4) In critical care, anesthesia, neonatal, and pediatric applications or other healthcare settings where use of the syringe infusion pump can be monitored or supervised by a clinician; 5) Inside the MRI room mounted outside the 150 Gauss line and with shielded magnets of field strength of 1.5 Tesla. recalled?

Yes. FDA published recall Z-1135-2018 on March 28, 2018. Reported hazard: Certain Medfusion Syringe Pump Models, Series 3100, 3500, and 4000, may not recognize or may misidentify loaded medication syringes. The inability of a pump to recognize a syringe (i.e. the size of the syringe is unknown to the pump) results in an inability to complete pump programming. Misidentification of a syringe is where the pump misinterprets the syringe size.

What should I do if I have this product?

Voluntary: Firm initiated · Letter