Ionolux, Catalog (REF) Numbers: a) 2115, b)2117, c) 2118, d) 2119, e) 2120, f) 2122; dental cement
Published February 28, 2024 · Recall initiated: January 25, 2024
Summary
FDA records medical device Class II recall Z-1143-2024, published February 28, 2024 and initiated January 25, 2024. Recalling firm: Voco GmbH; Status: Ongoing; Reported quantity: 1021 units; Lot: a) 2115, UDI/DI +E22121151, Lot Numbers: 2346779, 2401515; b) 2117, UDI/DI +E22121171, Lot Numbers: 2347832; c) 2118, UDI/DI +E22121181, Lot Numbers: 2349536; d) 2119, UDI/DI +E22121191, Lot Numbers: 2348501; e) 2120, UDI/DI +E22121201, Lot Numbers: 2350638; f) 2122, UDI/DI +E22121221, Lot Numbers: 2349527. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Due to an error in the manufacturing process, it is possible that an incorrect plunger has been fitted to some capsules. This plunger does not contain the correct reactive component for Ionolux. When these capsules are activated, curing is significantly delayed and incomplete. Processing according to the instructions for use is not possible.
What to do
Voluntary: Firm initiated · E-Mail
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Voco GmbH
- Distributor
- Worldwide distribution: US distribution to states of: IA, NY, PA, WA; and OUS (foreign) to countries of: Belgium, Bulgaria, Denmark, Germany, France, Greece, Great Britain, India, Italy, Canada, Quatar, Lithuania, Netherlands, Northern Ireland, Norway, Poland, Portugal, Romania, Sweden, Switzerland, Slovakia, Spain, Czech Republic, Australia, Saudi Arabia, & Ireland.
- Lot
- a) 2115, UDI/DI +E22121151, Lot Numbers: 2346779, 2401515; b) 2117, UDI/DI +E22121171, Lot Numbers: 2347832; c) 2118, UDI/DI +E22121181, Lot Numbers: 2349536; d) 2119, UDI/DI +E22121191, Lot Numbers: 2348501; e) 2120, UDI/DI +E22121201, Lot Numbers: 2350638; f) 2122, UDI/DI +E22121221, Lot Numbers: 2349527
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- Germany
Related recalls
- FDAIonoStar Plus, Catalog (REF) Numbers: a) 2540, b)2543, c) 2544, d) 2545, e) 2546, f) 2547, g) 2549; dentalFebruary 28, 2024
- FDAPhilips Allura Xper Interventional Fluoroscopic X-ray Systems, specifically the Disk Bay of the PC componentFebruary 28, 2024
- FDAAligned Medical Solutions, Closure Kit, AMS13043, STERILEEOFebruary 28, 2024
- FDAPhilips Azurion Interventional Fluoroscopic X-ray Systems, specifically the Disk Bay of the PC component ofFebruary 28, 2024
- FDAAligned Medical Solutions, In House Ocular Pack, AMS12947, STERILEEOFebruary 28, 2024
- FDAGrandio SO Heavy Flow Caps, Color A3 - Product Usage: Filling minimally invasive cavities of all classes -October 9, 2019
Questions about this recall
Was Ionolux, Catalog (REF) Numbers: a) 2115, b)2117, c) 2118, d) 2119, e) 2120, f) 2122; dental cement recalled?
Yes. FDA published recall Z-1143-2024 on February 28, 2024. Reported hazard: Due to an error in the manufacturing process, it is possible that an incorrect plunger has been fitted to some capsules. This plunger does not contain the correct reactive component for Ionolux. When these capsules are activated, curing is significantly delayed and incomplete. Processing according to the instructions for use is not possible.
What should I do if I have this product?
Voluntary: Firm initiated · E-Mail