FDAFDA · Food and drugs·Recall number Z-1144-2017

LIFEPAK 15 Monitor/Defibrillator with End-Tidal CO2 option. Intended for use by trained medical personnel in out-of-doors and indoor emergency care settings within the environmental conditions specified in the Operating Instructions.

Published February 15, 2017 · Recall initiated: January 16, 2017

Summary

FDA records medical device Class II recall Z-1144-2017, published February 15, 2017 and initiated January 16, 2017. Recalling firm: Physio-Control, Inc.; Status: Terminated; Reported quantity: 2584 units total (1501 units in the US; 1034 units international; and 49 units owned by Physio). 50 modules total; Lot: LOT NUMBERS OF AFFECTED MODULE: Lot # 1506, 1520, 1532, 1544, and 1551. SERIAL NUMBERS OF AFFECTED DEVICES: 37930990, 38046566, 38130282, 38200467, 38374121, 38544771, 38561025, 38567823, 38577667, 38763423, 38805573, 39229417, 39656903, 39749477, 39757723, 39757723, 39905189, 40007558, 40007616, 40008237, 40074226, 40246186, 40637428, 40639397, 40881488, 41119327, 41441205, 41734558, 41908077, 42036185, 42721982, 42759819, 43273957, 43981182, 44128955, 44143195, 44143356, 44143399, 44143438, 44143481, 44143616, 44143674, 44143708, 44143786, 44143811, 44143814, 44143891, 44143894, 44143984, 44146278, 44146304, 44146408, 44148345, 44148588, 44148641, 44150897, 44151007, 44151084, 44151151, 44151156, 44151224, 44151362, 44151498, 44151591, 44151697, 44151790, 44151809, 44151862, 44151920, 44151995, 44152133, 44152249, 44152345, 44152361, 44156527, 44157048, 44157124, 44157189, 44157404, 44157697, 44157711, 44158485, 44158543, 44158614, 44158707, 44158804, 44159096, 44159162, 44159337, 44159406, 44159406, 44159491, 44159555, 44159615, 44159786, 44160212, 44166262, 44166304, 44166470, 44166503, 44166514, 44166523, 44166536, 44166547, 44166631, 44171950, 44174110, 44174285, 44174322, 44174756, 44175011, 44175045, 44175070, 44175103, 44179307, 44179460, 44179499, 44179545, 44179576, 44179644, 44179750, 44180084, 44180141, 44180381, 44180443, 44180522, 44180584, 44180619, 44180789, 44180819, 44180917, 44182111, 44182155, 44185896, 44185922, 44185964, 44186134, 44186190, 44186249, 44186332, 44186361, 44186435, 44186555, 44186767, 44186789, 44186823, 44186871, 44186922, 44186961, 44187023, 44187363, 44187401, 44187447, 44187475, 44187739, 44187787, 44188141, 44195485, 44195502, 44195519, 44195593, 44195691, 44195713, 44195735, 44195749, 44195756, 44195861, 44195948, 44196025, 44196065, 44196096, 44196147, 44196198, 44196225, 44196266, 44196325, 44196414, 44196485, 44196832, 44196858, 44196898, 44196974, 44197026, 44197066, 44197146, 44197168, 44197212, 44197274, 44197295, 44197327, 44197348, 44197382, 44197397, 44197467, 44197511, 44198596, 44198678, 44198865, 44198906, 44198922, 44198961, 44198976, 44199088, 44199223, 44199386, 44199424, 44199478, 44199525, 44199806, 44199969, 44200010, 44200055, 44200186, 44200649, 44200724, 44200766, 44200807, 44200910, 44201002, 44201062, 44201120, 44201156, 44201243, 44201301, 44201342, 44201384, 44201466, 44201500, 44201551, 44203033, 44203091, 44203121, 44203172, 44203200, 44203234, 44203258, 44203316, 44203327, 44203472, 44203564, 44203654, 44203856, 44203886, 44203969, 44204052, 44204268, 44204540, 44204717, 44204782, 44204853, 44205142, 44205292, 44205329, 44205365, 44205404, 44205516, 44205592, 44205633, 44205705, 44205764, 44205875, 44205960, 44206055, 44206131, 44206166, 44206245, 44207820, 44207861, 44207882, 44208048, 44208105, 44208127, 44208152, 44208202, 44208320, 44208450, 44208494, 44209302, 44209524, 44209555, 44209797, 44209919, 44210007, 44210219, 44210271, 44210344, 44210396, 44210473, 44210493, 44210544, 44210619, 44210911, 44210940, 44210957, 44210984, 44211004, 44211021, 44211321, 44211355, 44211394, 44219958, 44220117, 44220142, 44220340, 44220358, 44220381, 44220403, 44220534, 44220569, 44220650, 44220734, 44220757, 44220782, 44220818, 44220856, 44220881, 44220919, 44220947, 44221011, 44221083, 44221152, 44221229, 44221426, 44221458, 44221503, 44222683, 44222703, 44222723, 44224099, 44224106, 44224120, 44224121, 44224139, 44224146, 44224176, 44224178, 44224194, 44224197, 44224243, 44224269, 44224285, 44224295, 44224306, 44224314, 44224315, 44224414, 44224419, 44224559, 44224635, 44224743, 44224804, 44224845, 44224979, 44225033, 44225060, 44225117, 44225180, 44225227, 44225242, 44225258, 44225272, 44225368, 44225610, 44225663, 44225706, 44225860, 44225939, 44226003, 44226063, 44226136, 44226183, 44226221, 44226231, 44226283, 44226384, 44226403, 44226417, 44226480, 44226493, 44226528, 44226564, 44226612, 44226624, 44226654, 44228007, 44228009, 44228037, 44228039, 4. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

The End-Tidal CO2 (EtCO2) reading can intermittently show a value of XXX after start-up or during device operation.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Physio-Control, Inc.
Distributor
Worldwide distribution. LifePak 12 distributed in US states of AL, GA, NY, SC, and in Australia. LifePak 15 distributed nationwide in the US including Puerto Rico and to ALGERIA, ARGENTINA, AUSTRALIA, AUSTRIA, BAHRAIN, BELGIUM, CANADA, CHILE, CZECH REPUBLIC, DENMARK, FINLAND, FRANCE, FRENCH GUIANA, GERMANY, GREECE, HUNGARY, ICELAND, IRELAND, ITALY, LITHUANIA, LUXEMBOURG, NETHERLANDS, NEW ZEALAND, NORWAY, POLAND, PORTUGAL, SAUDI ARABIA, SENEGAL, SERBIA, SLOVENIA, SOUTH AFRICA, SPAIN, SWITZERLAND, TAHITI, UNITED ARAB EMIRATES, AND UNITED KINGDOM.
Lot
LOT NUMBERS OF AFFECTED MODULE: Lot # 1506, 1520, 1532, 1544, and 1551. SERIAL NUMBERS OF AFFECTED DEVICES: 37930990, 38046566, 38130282, 38200467, 38374121, 38544771, 38561025, 38567823, 38577667, 38763423, 38805573, 39229417, 39656903, 39749477, 39757723, 39757723, 39905189, 40007558, 40007616, 40008237, 40074226, 40246186, 40637428, 40639397, 40881488, 41119327, 41441205, 41734558, 41908077, 42036185, 42721982, 42759819, 43273957, 43981182, 44128955, 44143195, 44143356, 44143399, 44143438, 44143481, 44143616, 44143674, 44143708, 44143786, 44143811, 44143814, 44143891, 44143894, 44143984, 44146278, 44146304, 44146408, 44148345, 44148588, 44148641, 44150897, 44151007, 44151084, 44151151, 44151156, 44151224, 44151362, 44151498, 44151591, 44151697, 44151790, 44151809, 44151862, 44151920, 44151995, 44152133, 44152249, 44152345, 44152361, 44156527, 44157048, 44157124, 44157189, 44157404, 44157697, 44157711, 44158485, 44158543, 44158614, 44158707, 44158804, 44159096, 44159162, 44159337, 44159406, 44159406, 44159491, 44159555, 44159615, 44159786, 44160212, 44166262, 44166304, 44166470, 44166503, 44166514, 44166523, 44166536, 44166547, 44166631, 44171950, 44174110, 44174285, 44174322, 44174756, 44175011, 44175045, 44175070, 44175103, 44179307, 44179460, 44179499, 44179545, 44179576, 44179644, 44179750, 44180084, 44180141, 44180381, 44180443, 44180522, 44180584, 44180619, 44180789, 44180819, 44180917, 44182111, 44182155, 44185896, 44185922, 44185964, 44186134, 44186190, 44186249, 44186332, 44186361, 44186435, 44186555, 44186767, 44186789, 44186823, 44186871, 44186922, 44186961, 44187023, 44187363, 44187401, 44187447, 44187475, 44187739, 44187787, 44188141, 44195485, 44195502, 44195519, 44195593, 44195691, 44195713, 44195735, 44195749, 44195756, 44195861, 44195948, 44196025, 44196065, 44196096, 44196147, 44196198, 44196225, 44196266, 44196325, 44196414, 44196485, 44196832, 44196858, 44196898, 44196974, 44197026, 44197066, 44197146, 44197168, 44197212, 44197274, 44197295, 44197327, 44197348, 44197382, 44197397, 44197467, 44197511, 44198596, 44198678, 44198865, 44198906, 44198922, 44198961, 44198976, 44199088, 44199223, 44199386, 44199424, 44199478, 44199525, 44199806, 44199969, 44200010, 44200055, 44200186, 44200649, 44200724, 44200766, 44200807, 44200910, 44201002, 44201062, 44201120, 44201156, 44201243, 44201301, 44201342, 44201384, 44201466, 44201500, 44201551, 44203033, 44203091, 44203121, 44203172, 44203200, 44203234, 44203258, 44203316, 44203327, 44203472, 44203564, 44203654, 44203856, 44203886, 44203969, 44204052, 44204268, 44204540, 44204717, 44204782, 44204853, 44205142, 44205292, 44205329, 44205365, 44205404, 44205516, 44205592, 44205633, 44205705, 44205764, 44205875, 44205960, 44206055, 44206131, 44206166, 44206245, 44207820, 44207861, 44207882, 44208048, 44208105, 44208127, 44208152, 44208202, 44208320, 44208450, 44208494, 44209302, 44209524, 44209555, 44209797, 44209919, 44210007, 44210219, 44210271, 44210344, 44210396, 44210473, 44210493, 44210544, 44210619, 44210911, 44210940, 44210957, 44210984, 44211004, 44211021, 44211321, 44211355, 44211394, 44219958, 44220117, 44220142, 44220340, 44220358, 44220381, 44220403, 44220534, 44220569, 44220650, 44220734, 44220757, 44220782, 44220818, 44220856, 44220881, 44220919, 44220947, 44221011, 44221083, 44221152, 44221229, 44221426, 44221458, 44221503, 44222683, 44222703, 44222723, 44224099, 44224106, 44224120, 44224121, 44224139, 44224146, 44224176, 44224178, 44224194, 44224197, 44224243, 44224269, 44224285, 44224295, 44224306, 44224314, 44224315, 44224414, 44224419, 44224559, 44224635, 44224743, 44224804, 44224845, 44224979, 44225033, 44225060, 44225117, 44225180, 44225227, 44225242, 44225258, 44225272, 44225368, 44225610, 44225663, 44225706, 44225860, 44225939, 44226003, 44226063, 44226136, 44226183, 44226221, 44226231, 44226283, 44226384, 44226403, 44226417, 44226480, 44226493, 44226528, 44226564, 44226612, 44226624, 44226654, 44228007, 44228009, 44228037, 44228039, 4
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was LIFEPAK 15 Monitor/Defibrillator with End-Tidal CO2 option. Intended for use by trained medical personnel in out-of-doors and indoor emergency care settings within the environmental conditions specified in the Operating Instructions. recalled?

Yes. FDA published recall Z-1144-2017 on February 15, 2017. Reported hazard: The End-Tidal CO2 (EtCO2) reading can intermittently show a value of XXX after start-up or during device operation.

What should I do if I have this product?

Voluntary: Firm initiated · Letter