EndoVive Replacement Button Kit w/ ENFIT 28F 2.8 CM, Outer Box UPN M00509500 or M00509510.
Published February 28, 2024 · Recall initiated: December 19, 2023
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Affected batches of the kit contain Stoma Measuring Devices sized to measure for an 18/24F replacement button, but are labeled as being for a 28F. Use of defective product may lead to undersizing the button, which may cause embedding with erosion into the gastric wall, tissue necrosis, infection, sepsis, and associate sequelae.
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- Boston Scientific Corporation
- Distributor
- International distribution to the country of Japan only.
- Lot
- UPN M00509500, GTIN (UDI-DI) 08714729902713, Lot 31073152; UPN M00509510, GTIN (UDI-DI) 08714729902720, Lots 31221751, 31761882, 32409573.
- Risk classification
- Class II