FDAFDA · Food and drugs·Recall number Z-1157-2013
APLIF Implants and Instruments, Part Numbers: 184008-10 Anterior Psoas Lumbar 18 x 40 x 08mm - 10degree 184010-10 Anterior Psoas Lumbar 18 x 40 x 10mm - 10degree 184012-10 Anterior Psoas Lumbar 18 x 40 x 12mm - 10degree 184014-10 Anterior Psoas Lumbar 18 x 40 x 14mm - 10degree 184016-10 Anterior Psoas Lumbar 18 x 40 x 16mm - 10degree 184008-00 Anterior Psoas Lumbar 18 x 40 x 08mm - 00degree 184010-00 Anterior Psoas Lumbar 18 x 40 x 10mm - 00degree 184012-00 Anterior Psoas Lumbar 18 x 40 x 12mm - 00degree 184014-00 Anterior Psoas Lumbar 18 x 40 x 14mm - 00degree 184016-00 Anterior Psoas Lumbar 18 x 40 x 16mm - 00degree 184510-08 Anterior Psoas Lumbar 18 x 45 x 08mm - 10degree 184510-10 Anterior Psoas Lumbar 18 x 45 x 10mm - 10degree 184512-10 Anterior Psoas Lumbar 18 x 45 x 12mm - 10degree 184514-10 Anterior Psoas Lumbar 18 x 45 x 14mm - 10degree 184516-10 Anterior Psoas Lumbar 18 x 45 x 16mm - 10degree 184510-00 Anterior Psoas Lumbar 18 x 45 x 10mm - 00degree 184512-00 Anterior Psoas Lumbar 18 x 45 x 12mm - 00degree 184514-00 Anterior Psoas Lumbar 18 x 45 x 14mm - 00degree 184516-00 Anterior Psoas Lumbar 18 x 45 x 16mm - 00degree 185008-10 Anterior Psoas Lumbar 18 x 50 x 08mm - 10degree 185010-10 Anterior Psoas Lumbar 18 x 50 x 10mm - 10degree 185012-10 Anterior Psoas Lumbar 18 x 50 x 12mm - 10degree 185014-10 Anterior Psoas Lumbar 18 x 50 x 14mm - 10degree 185016-10 Anterior Psoas Lumbar 18 x 50 x 16mm - 10degree 185008-00 Anterior Psoas Lumbar 18 x 50 x 08mm - 00degree 185010-00 Anterior Psoas Lumbar 18 x 50 x 10mm - 00degree 185012-00 Anterior Psoas Lumbar 18 x 50 x 12mm - 00degree 185014-00 Anterior Psoas Lumbar 18 x 50 x 14mm - 00degree 185016-00 Anterior Psoas Lumbar 18 x 50 x 16mm - 00degree 185508-10 Anterior Psoas Lumbar 18 x 55 x 08mm - 10degree 185510-10 Anterior Psoas Lumbar 18 x 55 x 10mm - 10degree 185512-10 Anterior Psoas Lumbar 18 x 55 x 12mm - 10degree 185514-10 Anterior Psoas Lumbar 18 x 55 x 14mm - 10degree 185516-10 Anterior Psoas Lumbar 18 x 55 x 16mm - 10degree 185508-00 Anterior Psoas Lumbar 18 x 55 x 08mm - 00degree 185510-00 Anterior Psoas Lumbar 18 x 55 x 10mm - 00degree 185512-00 Anterior Psoas Lumbar 18 x 55 x 12mm - 00degree 185514-00 Anterior Psoas Lumbar 18 x 55 x 14mm - 00degree 185516-00 Anterior Psoas Lumbar 18 x 55 x 16mm - 00degree The Eminent Spine Interbody Fusion System is indicated for intervertebral body fusion of the lumbar spine, from L2 to S1, in skeletally mature patients who have had six months of non-operative treatment.
Published May 1, 2013 · Recall initiated: March 8, 2013
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Spinal Solutions is recalling the APLIF system because it is not supported by adequate testing and documentation to demonstrate that it meets performance or safety standards. These inadequacies might result in product performance failures that could cause patient harm due to implant breakage, movement, or inadequate sterilization.
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- Spinal Solutions, LLC
- Distributor
- Nationwide Distribution including NV, WI, MD,and CA.
- Lot
- All lots are subject to this recall.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Sell products like this?
Recalert watches your whole catalog for recalls like this one, automatically.