Back to search
FDAFDA · Food and drugs·Recall number Z-1158-2023

Cardinal Health Presource Packs IV START TRAY I-Intended for IV access device CATALOG #: 01-5799A

Published March 1, 2023 · Recall initiated: November 18, 2022

Summary

FDA records medical device Class II recall Z-1158-2023, published March 1, 2023 and initiated November 18, 2022. Recalling firm: Cardinal Health 200, LLC; Status: Ongoing; Reported quantity: 36 EA; Lot: UDI-DI: 50195594381965(cs) 10195594381967(ea) Lot Number: 706265 Exp. Date: 5/1/2024. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Presource Kits contains the ICU MicroClave Clear Connector due to a manufacturing defect there is a gap between the connectors top and bottom housings which may lead to fluid leaks and loss of fluid path integrity

What to do

Voluntary: Firm initiated

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Cardinal Health 200, LLC
Distributor
US Nationwide distribution in the states of GA, PA.
Lot
UDI-DI: 50195594381965(cs) 10195594381967(ea) Lot Number: 706265 Exp. Date: 5/1/2024
Risk classification
Class II
Area (FDA)
device
Country
United States