Button-Style Ultrasound Transducer (Nautilus), 8-foot Cord 5700HAX, that may have been serviced with the following: Pacific Medical GE Corometrics Nautilus Ultrasound Board, NFCM9510; Pacific Medical GE Corometrics Nautilus Ultrasound Bottom Case, NFCM9515, Pacific Medical GE Corometrics Nautilus Ultrasound Cable Assembly, NFCM9520; Pacific Medical GE Corometrics Nautilus Ultrasound Crystal, NFCM9535
Published February 19, 2020 · Recall initiated: September 11, 2019
Summary
FDA records medical device Class II recall Z-1161-2020, published February 19, 2020 and initiated September 11, 2019. Recalling firm: Pacific Medical Group Inc.; Status: Ongoing; Reported quantity: 6481; Lot: All serial numbers serviced between 10/20/14 and 4/2/19. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Fetal/maternal monitor/monitoring systems serviced or remanufactured using non-OEM equivalent components that have not been appropriately verified or validated. Possible adverse clinical impact: inability to accurately detect and measure fetal heart rate, inability to accurately detect maternal peripheral oxygen saturation, inability to monitor uterine activity to follow contractions, inability to determine temperature, and maternal blood pressure, lack of electrocardiography tracing, patient/clinician burns, and delay in detecting maternal or fetal distress.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Pacific Medical Group Inc.
- Distributor
- Worldwide Distribution - US Nationwide in the states of::NH, NY, MN, PA, FL, AZ, LA, GA, IL, HI, CA, OR, SC, OH, NC, IA, OK, AL, MA, MD, WI, MI, AR, WA, SD, NV, KY, TX, MS, DE, NJ, NE, WV, MT, TN, VA, MO, VT, CO, CT, ME, WY, IN, UT, NM, AK, DC, ID, KS, ND, RI, NH. OUS: BEL, ARE, MEX, CAN, CYM, KOR, ZAF, BHR, ESP, ITA, FRA, VUT, AUS, GBR.
- Lot
- All serial numbers serviced between 10/20/14 and 4/2/19.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAUltrasound Transducer (M1356A), that may have been serviced with the following: Pacific Medical Philips M1356AFebruary 19, 2020
- FDALoop-Style Tocotransducer (Nautilus) 2264LAX, that may have been serviced with the following: Pacific MedicalFebruary 19, 2020
- FDAToco MP Transducer (M2734B), that may have been serviced with the following: Pacific Medical Philips M2734BFebruary 19, 2020
- FDAImplant Direct, CE 0086, REF: 855710, Legacy 3 Impant, Surface: SBM, Qty:1, Size:5.7mmD x 10mmL, 5.7mmDFebruary 19, 2020
- FDAROSA Brain 3.0 Application-BrainFebruary 19, 2020
- FDAPacific Medical GE Corometrics Nautilus Ultrasound Transducer w/wing 5700LAX, FFCM5260February 12, 2020
Questions about this recall
Was Button-Style Ultrasound Transducer (Nautilus), 8-foot Cord 5700HAX, that may have been serviced with the following: Pacific Medical GE Corometrics Nautilus Ultrasound Board, NFCM9510; Pacific Medical GE Corometrics Nautilus Ultrasound Bottom Case, NFCM9515, Pacific Medical GE Corometrics Nautilus Ultrasound Cable Assembly, NFCM9520; Pacific Medical GE Corometrics Nautilus Ultrasound Crystal, NFCM9535 recalled?
Yes. FDA published recall Z-1161-2020 on February 19, 2020. Reported hazard: Fetal/maternal monitor/monitoring systems serviced or remanufactured using non-OEM equivalent components that have not been appropriately verified or validated. Possible adverse clinical impact: inability to accurately detect and measure fetal heart rate, inability to accurately detect maternal peripheral oxygen saturation, inability to monitor uterine activity to follow contractions, inability to determine temperature, and maternal blood pressure, lack of electrocardiography tracing, patient/clinician burns, and delay in detecting maternal or fetal distress.
What should I do if I have this product?
Voluntary: Firm initiated · Letter