Oxford Partial Knee System Right Medial Tibial Trial Tray Size C Product Usage: Instruments for use with the Oxford Uni Partial Knee System
Published March 1, 2017 · Recall initiated: January 5, 2016
Summary
FDA records medical device Class II recall Z-1164-2017, published March 1, 2017 and initiated January 5, 2016. Recalling firm: Biomet U.K., Ltd.; Status: Terminated; Reported quantity: 72 units in total; Lot: Item: 32-420733, Lots: ZB160501, ZB160601. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Zimmer Biomet is conducting a medical device field action for various lot-specific Oxford Knee instruments due to potential alumina inclusions in the raw material batch used to produce the affected products. Inclusions contained within the finished product could lead to the cracking and separation of the instrument.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Biomet U.K., Ltd.
- Distributor
- Worldwide Distribution in the states of LA, NC, AR, MT, MO, CA, NY, KS, TX, MA, KY, OK, SC, NJ and the countries of Foreign: THAILAND, JAPAN, NETHERLANDS
- Lot
- Item: 32-420733, Lots: ZB160501, ZB160601
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United Kingdom
Related recalls
- FDAOxford Partial Knee System Phase 3 Tibial Template Right Medial Size B Product Usage: Instruments for use withMarch 1, 2017
- FDAOxford Partial Knee System Left Medial Tibial Trial Tray Size E Product Usage: Instruments for use with theMarch 1, 2017
- FDAOxford Partial Knee System Right Medial Tibial Trial Tray Size F Product Usage: Instruments for use with theMarch 1, 2017
- FDAOxford Partial Knee System Phase 3 Tibial Template Right Medial Size D Product Usage: Instruments for use withMarch 1, 2017
- FDAOxford Partial Knee System Right Medial Tibial Trial Tray Size C Oxford Partial Knee System Left Medial TibialMarch 1, 2017
- FDAOxford Partial Knee System Right Medial Tibial Trial Tray Size E Product Usage: Instruments for use with theMarch 1, 2017
Questions about this recall
Was Oxford Partial Knee System Right Medial Tibial Trial Tray Size C Product Usage: Instruments for use with the Oxford Uni Partial Knee System recalled?
Yes. FDA published recall Z-1164-2017 on March 1, 2017. Reported hazard: Zimmer Biomet is conducting a medical device field action for various lot-specific Oxford Knee instruments due to potential alumina inclusions in the raw material batch used to produce the affected products. Inclusions contained within the finished product could lead to the cracking and separation of the instrument.
What should I do if I have this product?
Voluntary: Firm initiated · Letter