FDAFDA · Food and drugs·Recall number Z-1177-2024

Azurion 5 with a Certeray generator-To perform image guidance in diagnostic, interventional, and minimally invasive surgery procedures Models: (1) 722227 (2) 722228

Published February 28, 2024 · Recall initiated: February 6, 2024

Summary

FDA records medical device Class II recall Z-1177-2024, published February 28, 2024 and initiated February 6, 2024. Recalling firm: PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.; Status: Ongoing; Reported quantity: 27 units; Lot: OUS (1) Model: 722227 SN: UDI: 75 (01)00884838099227(21)75 122 (01)00884838099227(21)122 131 (01)00884838099227(21)131 132 (01)00884838099227(21)132 165 (01)00884838099227(21)165 116 (01)00884838099227(21)116 136 (01)00884838099227(21)136.' (2) Model: 722228 Serial Number UDI 168 (01)00884838099234(21)168 197 (01)00884838099234(21)197 154 (01)00884838099234(21)154 193 (01)00884838099234(21)193 165 (01)00884838099234(21)165 163 (01)00884838099234(21)163 139 (01)00884838099234(21)139 202 (01)00884838099234(21)202 141 (01)00884838099234(21)141 200 (01)00884838099234(21)200 174 (01)00884838099234(21)174 178 (01)00884838099234(21)178 147 (01)00884838099234(21)147 208 (01)00884838099234(21)208 171 (01)00884838099234(21)171 167 (01)00884838099234(21)167 194 (01)00884838099234(21)194 176 (01)00884838099234(21)176 228 (01)00884838086777(10)2.0.0.10015 161 (01)00884838099234(21)161. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Generator may fail due to a potential short circuit in the Printed Circuit Board Assembly (PCBA) in PoInt EVR (Power Invertor). If a short circuit occurs, it will cause the fuses to trip and the system will become non-functional due to power loss, resulting in a potential delay of procedure or termination of procedure.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Distributor
Worldwide distribution - US Nationwide distribution in the states of NY, PA, PR, TN, TX, VA, WA and the countries of Algeria, Armenia, Austria, Brazil, China, Colombia, Croatia, Czech Republic, Ecuador, France, Germany, Hungary, India, Indonesia, Iraq, Israel, Italy, Japan, Kazakhstan, Korea, Republic of Kuwait, Lithuania, Macao, Macedonia, Malaysia, Morocco, Netherlands, Norway, Oman, Palestine, State of Panama, Poland, Portugal, Qatar, Romania, Russian Federation, Saudi Arabia, Saudi Arabia, Spain, Sweden, Switzerland, Taiwan, Tunisia, Turkey, Ukraine, United Kingdom.
Lot
OUS (1) Model: 722227 SN: UDI: 75 (01)00884838099227(21)75 122 (01)00884838099227(21)122 131 (01)00884838099227(21)131 132 (01)00884838099227(21)132 165 (01)00884838099227(21)165 116 (01)00884838099227(21)116 136 (01)00884838099227(21)136.' (2) Model: 722228 Serial Number UDI 168 (01)00884838099234(21)168 197 (01)00884838099234(21)197 154 (01)00884838099234(21)154 193 (01)00884838099234(21)193 165 (01)00884838099234(21)165 163 (01)00884838099234(21)163 139 (01)00884838099234(21)139 202 (01)00884838099234(21)202 141 (01)00884838099234(21)141 200 (01)00884838099234(21)200 174 (01)00884838099234(21)174 178 (01)00884838099234(21)178 147 (01)00884838099234(21)147 208 (01)00884838099234(21)208 171 (01)00884838099234(21)171 167 (01)00884838099234(21)167 194 (01)00884838099234(21)194 176 (01)00884838099234(21)176 228 (01)00884838086777(10)2.0.0.10015 161 (01)00884838099234(21)161
Risk classification
Class II
Area (FDA)
device
Country
Netherlands

Related recalls

Questions about this recall

Was Azurion 5 with a Certeray generator-To perform image guidance in diagnostic, interventional, and minimally invasive surgery procedures Models: (1) 722227 (2) 722228 recalled?

Yes. FDA published recall Z-1177-2024 on February 28, 2024. Reported hazard: Generator may fail due to a potential short circuit in the Printed Circuit Board Assembly (PCBA) in PoInt EVR (Power Invertor). If a short circuit occurs, it will cause the fuses to trip and the system will become non-functional due to power loss, resulting in a potential delay of procedure or termination of procedure.

What should I do if I have this product?

Voluntary: Firm initiated · Letter