Presource PBDS, Extremity, Kit, Circulator
Published May 15, 2013 · Recall initiated: March 26, 2013
Summary
FDA records medical device Class I recall Z-1198-2013, published May 15, 2013 and initiated March 26, 2013. Recalling firm: Cardinal Health, Medical Products & Services; Status: Terminated; Reported quantity: 4,571 Kits total; Lot: Catalog Number: PO24EXNPD01; Lots: 852429, 868825, 883022, 896400, 914524, 929984, 941696, 954317. Class I means a reasonable probability of serious adverse health consequences or death.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Various Presource Kits contain a pre-assembled Filter and Anesthesia Circuit that may contain outer plastic packaging on one or more components. If the packaging is removed without disassembling the components, remnants of the plastic from the packaging material may become lodged in the filter potentially causing an obstruction of airflow.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Cardinal Health, Medical Products & Services
- Distributor
- Distributed in the states of AR, MS, SC, and TN.
- Lot
- Catalog Number: PO24EXNPD01; Lots: 852429, 868825, 883022, 896400, 914524, 929984, 941696, 954317
- Risk classification
- Class I
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAPresource PBDS, Shoulder Arthroscopy, Kit, CirculatorMay 15, 2013
- FDAPresource PBDS, Knee Arthroscopy, Kit, CirculatorMay 15, 2013
- FDAPresource PBDS, Hand, Kit, CirculatorMay 15, 2013
- FDAPresource PBDS, Shoulder Procedure, Kit, CirculatorMay 15, 2013
- FDAPresource PBDS, Major Abdominal, Kit, CirculatorMay 15, 2013
- FDAPresource PBDS, General Laparoscopy, Kit, CirculatorMay 15, 2013
Questions about this recall
Was Presource PBDS, Extremity, Kit, Circulator recalled?
Yes. FDA published recall Z-1198-2013 on May 15, 2013. Reported hazard: Various Presource Kits contain a pre-assembled Filter and Anesthesia Circuit that may contain outer plastic packaging on one or more components. If the packaging is removed without disassembling the components, remnants of the plastic from the packaging material may become lodged in the filter potentially causing an obstruction of airflow.
What should I do if I have this product?
Voluntary: Firm initiated · Letter