FreeStyle Flash Blood Glucose Monitors System. Meter made in China; Distributed by TheraSense, Alameda, CA The FreeStyle Flash meter should only be used with FreeStyle Test Strips and FreeStyle Control Solution. Using other brands of test strips and control solutions with the FreeStyle Meter can product inaccurate results. Intended to monitor blood glucose from samples taken from the body. IVD use only,
Published March 26, 2014 · Recall initiated: February 19, 2014
Summary
FDA records medical device Class I recall Z-1212-2014, published March 26, 2014 and initiated February 19, 2014. Recalling firm: Abbott Diabetes Care, Inc.; Status: Terminated; Reported quantity: 25,203 units; Lot: Catalog Number 17002 NDC 99073-0170-02. All serial numbers. Class I means a reasonable probability of serious adverse health consequences or death.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Abbott Diabetes Care has identified through internal testing and investigation that all non-applied voltage legacy meters have the potential to produce out of range control solution results and/or erroneously low blood glucose results when used in conjunction with Free Style test strips lots within expiry.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Abbott Diabetes Care, Inc.
- Distributor
- Worldwide Distribution: US (nationwide) and countries of: Austria, Canada, Germany, Israel, Netherlands, Norway, Sweden, Switzerland, United Kingdom.
- Lot
- Catalog Number 17002 NDC 99073-0170-02. All serial numbers
- Risk classification
- Class I
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAFreeStyle Libre 3 Sensors, REF: 72081-01, component of the FreeStyle Libre 3 Continuous Glucose MonitoringSeptember 4, 2024
- FDAFreeStyle Libre 3 App, part of Continuous Glucose Monitoring System, when used on the Android 13 OperatingJanuary 24, 2024
- FDAPrecision Xtra Blood Glucose & Ketone Monitoring System, UDI: 00093815715023, 00093815988144April 1, 2020
- FDAFreeStyle Blood Glucose Monitors System. For in vitro diagnostic use only. Store at room temperature below 86March 26, 2014
- FDAMar Cor Purification Millenium HX Portable Reverse Osmosis System. Model 3025975. Product Usage: WaterMarch 26, 2014
- FDADeRoyal(R) LIMB HOLDERS, COTTON ADJUSTABLE, UNIVERSAL, REF M2017 Product Usage: Limb holders are intended forMarch 26, 2014
Questions about this recall
Was FreeStyle Flash Blood Glucose Monitors System. Meter made in China; Distributed by TheraSense, Alameda, CA The FreeStyle Flash meter should only be used with FreeStyle Test Strips and FreeStyle Control Solution. Using other brands of test strips and control solutions with the FreeStyle Meter can product inaccurate results. Intended to monitor blood glucose from samples taken from the body. IVD use only, recalled?
Yes. FDA published recall Z-1212-2014 on March 26, 2014. Reported hazard: Abbott Diabetes Care has identified through internal testing and investigation that all non-applied voltage legacy meters have the potential to produce out of range control solution results and/or erroneously low blood glucose results when used in conjunction with Free Style test strips lots within expiry.
What should I do if I have this product?
Voluntary: Firm initiated · Letter