The Enzyme 2 Calibrator is an in vitro diagnostic product for the calibration of Alanine Aminotransferase (ALTI and ALTI) and Aspartate Aminotransferase (AST) methods on the Dimension Vista System. For prescription use only.

Published February 22, 2017 · Recall initiated: May 5, 2016

Summary

FDA records medical device Class III recall Z-1216-2017, published February 22, 2017 and initiated May 5, 2016. Recalling firm: Siemens Healthcare Diagnostics, Inc.; Status: Terminated; Reported quantity: 13,147 units; Lot: Lot Number: 4MD048, 5DD026, 5ED071, 5GD004, 5JD058, 5LD061. Class III means the product is not likely to cause adverse health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Complaints of positive shift of patient and QC results when the ALTI assay is calibrated with the Dimension Vista Enzyme 2 calibrator. Internal investigation has confirmed that patient samples and QC material demonstrate a shift of up to 10% in comparison to results obtained using earlier calibrator lots.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Siemens Healthcare Diagnostics, Inc.
Distributor
Distributed throughout the United States
Lot
Lot Number: 4MD048, 5DD026, 5ED071, 5GD004, 5JD058, 5LD061
Risk classification
Class III
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was The Enzyme 2 Calibrator is an in vitro diagnostic product for the calibration of Alanine Aminotransferase (ALTI and ALTI) and Aspartate Aminotransferase (AST) methods on the Dimension Vista System. For prescription use only. recalled?

Yes. FDA published recall Z-1216-2017 on February 22, 2017. Reported hazard: Complaints of positive shift of patient and QC results when the ALTI assay is calibrated with the Dimension Vista Enzyme 2 calibrator. Internal investigation has confirmed that patient samples and QC material demonstrate a shift of up to 10% in comparison to results obtained using earlier calibrator lots.

What should I do if I have this product?

Voluntary: Firm initiated · Letter