Siemens Atellica IM aTG (100 test kit) Siemens Material Number (SMN): 10995461
Published May 8, 2019 · Recall initiated: December 7, 2018
Summary
FDA records medical device Class II recall Z-1235-2019, published May 8, 2019 and initiated December 7, 2018. Recalling firm: Siemens Healthcare Diagnostics, Inc; Status: Terminated; Reported quantity: 1489 kits; Lot: Lot Number/Exp. Date: kit lots ending in 317 and lower 72259303 12/13/2018 87903305 1/12/2019 94908307 2/11/2019 96439307 2/11/2019 16824309 3/14/2019 25822309 3/14/2019 28942311 4/27/2019 55978313 5/27/2019 70389317 6/29/2019. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Positive bias anti-thyroglobulin with ADVIA Centaur aTG kit lots ending in 316 and lower and Atellica IM aTG kit lots ending in 317 and lower when compared to the standardization to World Health Organization (WHO) Reference Preparation MRC 65/93 stated in the Instructions for Use (IFU).
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Siemens Healthcare Diagnostics, Inc
- Distributor
- Nationwide Foreign: Belgium®¿¿¿¿¿¿¿ Croatia®¿¿¿¿¿¿¿ Czech®Republic® Denmark®¿¿¿¿¿¿¿ France®¿¿¿¿¿¿¿¿ Germany®¿¿¿¿¿¿¿ Greece®¿¿¿¿¿¿¿¿ Ireland®¿¿¿¿¿¿¿ Italy®¿¿¿¿¿¿¿¿¿ Poland®¿¿¿¿¿¿¿¿ Portugal®¿¿¿¿¿¿ Slovakia®¿¿¿¿¿¿ Spain®¿¿¿¿¿¿¿¿¿ Switzerland®¿¿¿ Turkey®¿¿¿¿¿¿ BOGOTA QUARRY BAY, HONG KONG YANGON SANTIAGO DE CHILE MONTEVIDEO BUENOS AIRES LIMA JOINVILLE, SC, BRAZIL Selangor Darul Ehsan Dhaka Ha Noi City Gurgaon, Haryana Hanoi-Hoangmai ASUNCION SHANGHAI Seoul 120 SINGAPORE RIYADH 11411 TAIPEI R.O.C. QUITO HUNTINGWOOD, NEW SOUTH WALES ZAPOPAN Ho Chi Minh-District 3 Huaykwang, Bangkok RIYADH SAN JOSE CHIBA LAVAL HO CHI MINH CITY MANAGUA Gurgaon, Haryana
- Lot
- Lot Number/Exp. Date: kit lots ending in 317 and lower 72259303 12/13/2018 87903305 1/12/2019 94908307 2/11/2019 96439307 2/11/2019 16824309 3/14/2019 25822309 3/14/2019 28942311 4/27/2019 55978313 5/27/2019 70389317 6/29/2019
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDASiemens ADVIA Centaur Folate 100 test kit-for IVD of folate in serum or red blood cells SMN:10310308January 29, 2020
- FDASiemensADVIA Centaur aTG (500 test kit) Siemens Material Number (SMN): 10492399May 8, 2019
- FDAAtellica IM aTG (500 test kit) Siemens Material Number (SMN): 10995462May 8, 2019
- FDAStanganellis Italian Foods, Pumpkin Cake Roll, Net Wt. 24 oz. . One pie per package . Box.May 8, 2019
- FDAAutoDELFIA® hCG Kit, In Vitro Diagnostic for the determination of human gonadotropin (hCG) Product Code:May 8, 2019
- FDAAtellica IM CKMB Calibrator, Cat No. 10995532 Product Usage: For in vitro diagnostic use for evaluating theOctober 24, 2018
Questions about this recall
Was Siemens Atellica IM aTG (100 test kit) Siemens Material Number (SMN): 10995461 recalled?
Yes. FDA published recall Z-1235-2019 on May 8, 2019. Reported hazard: Positive bias anti-thyroglobulin with ADVIA Centaur aTG kit lots ending in 316 and lower and Atellica IM aTG kit lots ending in 317 and lower when compared to the standardization to World Health Organization (WHO) Reference Preparation MRC 65/93 stated in the Instructions for Use (IFU).
What should I do if I have this product?
Voluntary: Firm initiated · Letter