FreeStyle InsuLinx Blood Glucose Monitoring System: For In Vitro Diagnostic Use only. Use only with FreeStyle InsuLinx test strips. Abbott Diabetes Care, Inc. 1360 South Loop Road Alameda, CA 94502 USA The FreeStyle InsuLinx Blood Glucose Test Strips are for use with the FreeStyle InsuLinx Blood Glucose Meter to quantitatively measure glucose in capillary whole blood samples drawn from the fingertip.
Published May 15, 2013 · Recall initiated: April 15, 2013
Summary
FDA records medical device Class I recall Z-1245-2013, published May 15, 2013 and initiated April 15, 2013. Recalling firm: Abbott Diabetes Care, Inc.; Status: Terminated; Reported quantity: 335,273 meters; 209,336 remain at firm.; Lot: Part Numbers: 71142-70 71143-70 71145-70 71150-70: All serial and lot numbers. Class I means a reasonable probability of serious adverse health consequences or death.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
AT rare, extremely high glucose levels (1024 mg/dL and above), the FreeStyle InsuLinx Blood Glucose Meter may provide an inaccurate reading.
What to do
Voluntary: Firm initiated · Press Release
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Abbott Diabetes Care, Inc.
- Distributor
- Nationwide Distribution
- Lot
- Part Numbers: 71142-70 71143-70 71145-70 71150-70: All serial and lot numbers.
- Risk classification
- Class I
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAFreeStyle Libre 3 Sensors, REF: 72081-01, component of the FreeStyle Libre 3 Continuous Glucose MonitoringSeptember 4, 2024
- FDAFreeStyle Libre 3 App, part of Continuous Glucose Monitoring System, when used on the Android 13 OperatingJanuary 24, 2024
- FDAPrecision Xtra Blood Glucose & Ketone Monitoring System, UDI: 00093815715023, 00093815988144April 1, 2020
- FDAFreeStyle Blood Glucose Monitors System. For in vitro diagnostic use only. Store at room temperature below 86March 26, 2014
- FDAWallace Sure-Pro® Single Stage Embryo Replacement Catheter Product Code: PP623 Sure-Pro® Embryo ReplacementMay 15, 2013
- FDAMercury Spinal System DIA 5.5 x 30 MM Lordosed Rod, Part Number: 60100-030. Intended to provide immobilizationMay 15, 2013
Questions about this recall
Was FreeStyle InsuLinx Blood Glucose Monitoring System: For In Vitro Diagnostic Use only. Use only with FreeStyle InsuLinx test strips. Abbott Diabetes Care, Inc. 1360 South Loop Road Alameda, CA 94502 USA The FreeStyle InsuLinx Blood Glucose Test Strips are for use with the FreeStyle InsuLinx Blood Glucose Meter to quantitatively measure glucose in capillary whole blood samples drawn from the fingertip. recalled?
Yes. FDA published recall Z-1245-2013 on May 15, 2013. Reported hazard: AT rare, extremely high glucose levels (1024 mg/dL and above), the FreeStyle InsuLinx Blood Glucose Meter may provide an inaccurate reading.
What should I do if I have this product?
Voluntary: Firm initiated · Press Release