FreeStyle InsuLinx Blood Glucose Monitoring System: For In Vitro Diagnostic Use only. Use only with FreeStyle InsuLinx test strips. Abbott Diabetes Care, Inc. 1360 South Loop Road Alameda, CA 94502 USA The FreeStyle InsuLinx Blood Glucose Test Strips are for use with the FreeStyle InsuLinx Blood Glucose Meter to quantitatively measure glucose in capillary whole blood samples drawn from the fingertip.

Published May 15, 2013 · Recall initiated: April 15, 2013

Summary

FDA records medical device Class I recall Z-1245-2013, published May 15, 2013 and initiated April 15, 2013. Recalling firm: Abbott Diabetes Care, Inc.; Status: Terminated; Reported quantity: 335,273 meters; 209,336 remain at firm.; Lot: Part Numbers: 71142-70 71143-70 71145-70 71150-70: All serial and lot numbers. Class I means a reasonable probability of serious adverse health consequences or death.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

AT rare, extremely high glucose levels (1024 mg/dL and above), the FreeStyle InsuLinx Blood Glucose Meter may provide an inaccurate reading.

What to do

Voluntary: Firm initiated · Press Release

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Abbott Diabetes Care, Inc.
Distributor
Nationwide Distribution
Lot
Part Numbers: 71142-70 71143-70 71145-70 71150-70: All serial and lot numbers.
Risk classification
Class I
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was FreeStyle InsuLinx Blood Glucose Monitoring System: For In Vitro Diagnostic Use only. Use only with FreeStyle InsuLinx test strips. Abbott Diabetes Care, Inc. 1360 South Loop Road Alameda, CA 94502 USA The FreeStyle InsuLinx Blood Glucose Test Strips are for use with the FreeStyle InsuLinx Blood Glucose Meter to quantitatively measure glucose in capillary whole blood samples drawn from the fingertip. recalled?