Back to search
FDAFDA · Food and drugs·Recall number Z-1245-2017

Keyspan High-High Speed USB to Serial Adapter Product Usage: The Reporting Software is an application which serves as middle ware between the Tosoh Automated HPLC G8 analyzer and an LIS or as a stand alone data repository for the analyzer. It is designed to store and manage the received data and make reports as necessary. The application is installed on Dell All- in-One computers that do not have serial (RS232) ports only USB ports. The G8 analyzer does not have USB ports. The Key Span Adaptor connects the serial port on the G8 to the USB port on the Dell computer. Once installed the Keyspan adapter is assigned a COM port. The default for the firmware is dynamic which randomly assigns a port for the transmission of data. The port assignment can change when the power for the analyzer or computer is shut down or the cable is physically disconnected and then reconnected. The default can be changed to geographic which allows the port assignment to be static regardless of a power outage.

Published March 1, 2017 · Recall initiated: December 22, 2016

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Power outages causes reporting software to shutdown.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Tosoh Smd Inc
Distributor
US Nationwide Distribution in the states of CT, MA, ME, MN, MO, NC, NE, NH, PA, SD, TX, VA, & WV.
Lot
All G8's with Reporting Software interfaced with LIS.
Risk classification
Class II
Area (FDA)
device
Country
United States

Sell products like this?

Recalert watches your whole catalog for recalls like this one, automatically.