Back to search
FDAFDA · Food and drugs·Recall number Z-1248-2014

Gender Solutions" Natural-Knee® Flex System: N-K FLEX MBACK SZ 0 PAT N-K FLEX MBACK SZ 1 PAT N-K FLEX MBACK SZ 2 PAT

Published March 26, 2014 · Recall initiated: February 25, 2014

Summary

FDA records medical device Class II recall Z-1248-2014, published March 26, 2014 and initiated February 25, 2014. Recalling firm: Zimmer, Inc.; Status: Terminated; Reported quantity: 3638 units; Lot: Item Numbers 00541800000 00541800001 00541800002. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

A complaint was received for a discolored All Poly Tibia. Root cause analysis found that the discoloration was the result of free radical oxidation. The oxidation process was advanced prematurely as a result of the machined component being stored in a transparent container exposed to artificial light for six years prior to inert packaging and sterilization of the device.

What to do

Voluntary: Firm initiated

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Zimmer, Inc.
Distributor
Worldwide Distribution-USA (nationwide) including DC and the states of AK, AL, AR, AZ, CA, CO, CT, FL, IL, IN, KS, LA, ME, MI, MO, MS, NC, ND, NE, NJ, NY, OH, OK, OR, PA, TN, TX, UT, VA, WA, and WI, and the countries of Canada, Australia, India, Austria, Japan, Czech Republic, Seoul Korea, France, Venezuela, Germany, Luxembourg, Spain, and Switzerland.
Lot
Item Numbers 00541800000 00541800001 00541800002
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Gender Solutions" Natural-Knee® Flex System: N-K FLEX MBACK SZ 0 PAT N-K FLEX MBACK SZ 1 PAT N-K FLEX MBACK SZ 2 PAT recalled?

Yes. FDA published recall Z-1248-2014 on March 26, 2014. Reported hazard: A complaint was received for a discolored All Poly Tibia. Root cause analysis found that the discoloration was the result of free radical oxidation. The oxidation process was advanced prematurely as a result of the machined component being stored in a transparent container exposed to artificial light for six years prior to inert packaging and sterilization of the device.

What should I do if I have this product?

Voluntary: Firm initiated