MOST Options" System (MOST) HINGED TIB INS SZ 1 16MM HINGED TIB INS SZ 4 6MM M HINGED TIB INS SZ 4 9MM M HINGED TIB INS SZ 4 16MM MOST PROX TIB INS 6MM
Published March 26, 2014 · Recall initiated: February 25, 2014
Summary
FDA records medical device Class II recall Z-1250-2014, published March 26, 2014 and initiated February 25, 2014. Recalling firm: Zimmer, Inc.; Status: Terminated; Reported quantity: 3638 units; Lot: Item Number 502001016 502004006 502004009 502004016 502010006. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
A complaint was received for a discolored All Poly Tibia. Root cause analysis found that the discoloration was the result of free radical oxidation. The oxidation process was advanced prematurely as a result of the machined component being stored in a transparent container exposed to artificial light for six years prior to inert packaging and sterilization of the device.
What to do
Voluntary: Firm initiated
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Zimmer, Inc.
- Distributor
- Worldwide Distribution-USA (nationwide) including DC and the states of AK, AL, AR, AZ, CA, CO, CT, FL, IL, IN, KS, LA, ME, MI, MO, MS, NC, ND, NE, NJ, NY, OH, OK, OR, PA, TN, TX, UT, VA, WA, and WI, and the countries of Canada, Australia, India, Austria, Japan, Czech Republic, Seoul Korea, France, Venezuela, Germany, Luxembourg, Spain, and Switzerland.
- Lot
- Item Number 502001016 502004006 502004009 502004016 502010006
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAOff-Axis Comprehensive Shoulder System Small, Augment, Off-Axis, Reamer Guide, Model/Catalog Number: 110040240July 1, 2026
- FDABrand Name: Persona Revision Product Name: Persona Revision® Trabecular Metal® Femoral Distal AugmentJuly 1, 2026
- FDAOff-Axis Alliance Glenoid, Left, 4-Peg, Augment, Reamer Guide, Model/Catalog Number: 110040444July 1, 2026
- FDAOff-Axis Comprehensive Shoulder System, Medium, Augment, Off-Axis, Reamer Guide Model/Catalog Number:July 1, 2026
- FDANatural-Knee® N-K II DOMED ALL-POLY PAT, SZ0 N-K II DOMED ALL-POLY PAT, SZ1March 26, 2014
- FDANatural-Knee® II - Unicompartmental Knee Prosthesis: CONG TIB INS SZ1/2 9MM UN CONG TIB INS SZ1/2 11MM U CONGMarch 26, 2014
Questions about this recall
Was MOST Options" System (MOST) HINGED TIB INS SZ 1 16MM HINGED TIB INS SZ 4 6MM M HINGED TIB INS SZ 4 9MM M HINGED TIB INS SZ 4 16MM MOST PROX TIB INS 6MM recalled?
Yes. FDA published recall Z-1250-2014 on March 26, 2014. Reported hazard: A complaint was received for a discolored All Poly Tibia. Root cause analysis found that the discoloration was the result of free radical oxidation. The oxidation process was advanced prematurely as a result of the machined component being stored in a transparent container exposed to artificial light for six years prior to inert packaging and sterilization of the device.
What should I do if I have this product?
Voluntary: Firm initiated