Vantage Titan, Model: MRT-1504
Published March 13, 2024 · Recall initiated: January 29, 2024
Summary
FDA records medical device Class II recall Z-1250-2024, published March 13, 2024 and initiated January 29, 2024. Recalling firm: Canon Medical System, USA, INC.; Status: Ongoing; Reported quantity: 40; Lot: UDI-DI: 04987670102400. Serial Numbers: U4B15Y2102, U5B15Z2001, U4B1492080, U4B1582100, U4B1582099, U4B1522091, U4B1512089, U4B1432067, U4B1522093, U4B1452073, U4B1442069, U4B1442068, U4B1642106, U4B1432066, U4B15X2101, U4B14Y2083, U4B1452071, U4B1512087, U4B1652107, U4B1552097, U4B1482078, U4B1512090, U4B1552098, U4B1442070, U4B14Y2082, U4B1482076, U4B14Z2085, U4B14Z2086, U4B1512088, U4B1612103, U4B1472075, U4B1422065, U4B1462074, U4B14Y2084, U4B1632105, U4B1452072, U4B1492081, U4B1482079, U4B1612104, U4B1482077. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
For some MRI systems, it has been found that some of the maximum Spatial Field Gradient (SFG) values displayed in "System Information" of the operation window and some of the maximum SFG values described in the safety manual are lower than the correct values, which may cause a patient with an MR conditional device to feel some discomfort during the MR scan.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Canon Medical System, USA, INC.
- Distributor
- Worldwide - US Nationwide distribution including in the states of MD, IL, WV, CA, AZ, PR, CO, GA, PA, WA, UT, OK, KY, IA, IN, TN, CT, WI, OH, AL, MT, NM, TX, MO, MN, OR, ID, WY, NV, AR, FL, SD, MI, NY, NJ, KS, LA, NH, VT, MS, SC, ME and the country of DO.
- Lot
- UDI-DI: 04987670102400. Serial Numbers: U4B15Y2102, U5B15Z2001, U4B1492080, U4B1582100, U4B1582099, U4B1522091, U4B1512089, U4B1432067, U4B1522093, U4B1452073, U4B1442069, U4B1442068, U4B1642106, U4B1432066, U4B15X2101, U4B14Y2083, U4B1452071, U4B1512087, U4B1652107, U4B1552097, U4B1482078, U4B1512090, U4B1552098, U4B1442070, U4B14Y2082, U4B1482076, U4B14Z2085, U4B14Z2086, U4B1512088, U4B1612103, U4B1472075, U4B1422065, U4B1462074, U4B14Y2084, U4B1632105, U4B1452072, U4B1492081, U4B1482079, U4B1612104, U4B1482077
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAALPHENIX INTERVENTIONAL ANGIOGRAPHY SYSTEM REF: INFX-8000FNovember 4, 2026
- FDAALPHENIX INTERVENTIONAL ANGIOGRAPHY SYSTEM REF: INFX-8000CNovember 4, 2026
- FDAALPHENIX INTERVENTIONAL ANGIOGRAPHY SYSTEM REF: INFX-8000VNovember 4, 2026
- FDAAlphenix INFX-8000C/B5October 14, 2026
- FDAMRI system: Vantage Elan, Model: MRT-2020March 13, 2024
- FDAInvacare Homecare Series Bed and Invacare G-Series Beds Containing Component Model Numbers: (1) G50 (2) G53March 13, 2024
Questions about this recall
Was Vantage Titan, Model: MRT-1504 recalled?
Yes. FDA published recall Z-1250-2024 on March 13, 2024. Reported hazard: For some MRI systems, it has been found that some of the maximum Spatial Field Gradient (SFG) values displayed in "System Information" of the operation window and some of the maximum SFG values described in the safety manual are lower than the correct values, which may cause a patient with an MR conditional device to feel some discomfort during the MR scan.
What should I do if I have this product?
Voluntary: Firm initiated · Letter