Back to search
FDAFDA · Food and drugs·Recall number Z-1254-2014

Natural-Knee¿ Congruent Tibial Insert with Screw: SF-CON TIB INS LT SZ1/2 1 SF-CON TIB INS RT SZ1/2 9 SF-CON TIB INS RT SZ3/4/5

Published March 26, 2014 · Recall initiated: February 25, 2014

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

A complaint was received for a discolored All Poly Tibia. Root cause analysis found that the discoloration was the result of free radical oxidation. The oxidation process was advanced prematurely as a result of the machined component being stored in a transparent container exposed to artificial light for six years prior to inert packaging and sterilization of the device.

What to do

Voluntary: Firm initiated

Recall details

Manufacturer
Zimmer, Inc.
Distributor
Worldwide Distribution-USA (nationwide) including DC and the states of AK, AL, AR, AZ, CA, CO, CT, FL, IL, IN, KS, LA, ME, MI, MO, MS, NC, ND, NE, NJ, NY, OH, OK, OR, PA, TN, TX, UT, VA, WA, and WI, and the countries of Canada, Australia, India, Austria, Japan, Czech Republic, Seoul Korea, France, Venezuela, Germany, Luxembourg, Spain, and Switzerland.
Lot
Item Number 620005819 620005909 620006916
Risk classification
Class II
Area (FDA)
device
Country
United States