Altivate Anatomic(TM) INSTRUMENTATION, PEGGED GLENOID TRIAL, SIZE 50mm, REF 804-07-500 Product Usage: The AltiVate Anatomic Shoulder System is a total shoulder system indicated as an anatomic shoulder joint replacement for patients suffering from pain and dysfunction.
Published April 4, 2018 · Recall initiated: January 29, 2018
Summary
FDA records medical device Class II recall Z-1259-2018, published April 4, 2018 and initiated January 29, 2018. Recalling firm: Encore Medical, Lp; Status: Completed; Reported quantity: 92 units; Lot: All Serial Numbers. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Completed — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
It was discovered during a surgery that the design of the radius at the base of the pegs of the Anatomic Glenoid Trial (Part Numbers: 804-07-380, 804-07-420, 804-07-460, 804-07-500, and 804-07-540) is much larger on the trails than on the implants
What to do
Voluntary: Firm initiated · E-Mail
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Encore Medical, Lp
- Distributor
- US Nationwide Distribution
- Lot
- All Serial Numbers
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAAltivate Anatomic(TM) INSTRUMENTATION, PEGGED GLENOID TRIAL, SIZE 54mm, REF 804-07-540 Product Usage: TheApril 4, 2018
- FDARevolution CMS w/ Femoral Breakaway Nozzle & Med. Press., Product Number: 0606-573-000, UDI: (01)4546540912114April 4, 2018
- FDAPCD Kit: Long 90, 13g Match-Ground w/ Bevel, Product Number: 0506-485-000, UDI: (01)4546540907035 for spinalApril 4, 2018
- FDACobalt HV with Gentamicin, Cobalt Bone Cement 40GM, REF 402283, QTY 1, STERILE, Rx Only, Mfgr: BIOMETNovember 15, 2017
- FDACobalt MV Bone Cement 40GM, REF 402438, QTY 1, STERILE, Rx Only, Mfgr: BIOMET ORTHOPEDICS Cobalt MV and HVNovember 15, 2017
- FDACobalt MV with Gentamicin, Cobalt Bone Cement 40GM, REF 402439, QTY 1, STERILE, Rx Only, Mfgr: BIOMETNovember 15, 2017
Questions about this recall
Was Altivate Anatomic(TM) INSTRUMENTATION, PEGGED GLENOID TRIAL, SIZE 50mm, REF 804-07-500 Product Usage: The AltiVate Anatomic Shoulder System is a total shoulder system indicated as an anatomic shoulder joint replacement for patients suffering from pain and dysfunction. recalled?
Yes. FDA published recall Z-1259-2018 on April 4, 2018. Reported hazard: It was discovered during a surgery that the design of the radius at the base of the pegs of the Anatomic Glenoid Trial (Part Numbers: 804-07-380, 804-07-420, 804-07-460, 804-07-500, and 804-07-540) is much larger on the trails than on the implants
What should I do if I have this product?
Voluntary: Firm initiated · E-Mail