Altivate Anatomic(TM) INSTRUMENTATION, PEGGED GLENOID TRIAL, SIZE 50mm, REF 804-07-500 Product Usage: The AltiVate Anatomic Shoulder System is a total shoulder system indicated as an anatomic shoulder joint replacement for patients suffering from pain and dysfunction.
Published April 4, 2018 · Recall initiated: January 29, 2018
Status when published: Completed — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
It was discovered during a surgery that the design of the radius at the base of the pegs of the Anatomic Glenoid Trial (Part Numbers: 804-07-380, 804-07-420, 804-07-460, 804-07-500, and 804-07-540) is much larger on the trails than on the implants
What to do
Voluntary: Firm initiated · E-Mail
Recall details
- Manufacturer
- Encore Medical, Lp
- Distributor
- US Nationwide Distribution
- Lot
- All Serial Numbers
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States