Halogen Lamp component in RetCam 3 System - Clarity Medical Systems, Inc; Pleasanton, CA. General ophthalmic imaging including retinal, corneal and external imaging. Photo-documentation of pediatric ocular diseases including retinopathy of prematurity. Screening for Type-2 re-threshold retinopathy of prematurity or treatment requiring ROP.
Published March 26, 2014 · Recall initiated: March 7, 2014
Summary
FDA records medical device Class II recall Z-1265-2014, published March 26, 2014 and initiated March 7, 2014. Recalling firm: Clarity Medical Systems Inc; Status: Terminated; Reported quantity: 234 lamps; Lot: Lot of Halogan Lamp Component is #020612C. RetCam 3, Part number 21-100500. RetCam 3 affected serial numbers: RC3066 RC3079 RC6013 RC6033 RC6034 RC6035 RC6036 RC6037 RC6038 RC6039 RC6040 RC6041 RC6043 RC6044 RC6045 RC6047 RC6049 RC6050 RC6053 RC6054 RC6055 RC6056 RC6057 RC6058 RC6059 RC6060 RC6063 RC6064 RC6066 RC6067 RC6068 RC6069 RC6070 RC6071 RC6072 RC6073 RC6074 RC6075 RC6076 RC6077 RC6078 RC6079 RC6080 RC6081 RC6082 RC6083 RC6084 RC6085 RC6086 RC6087 RC6088 RC6089 RC6090 RC6092 RC6093 RC6094 RC6095 RC6096 RC6098 RC6099 RC6101 RC6102 RC6103 RC6104 RC6105 RC6106 RC6107 RC6108 RC6109 RC6110 RC6111 RC6112 RC6113 RC6301 RC6302 RC6303 RC6304 RC6305 RC6306 RC6307 RC6309 RC6310 RC6311 RC6312 RC6313 RC6314 RC6315 RC6316 RC6317 RC6318 RC6319 RC6320 RC6321 RC6322 RC6323 RC6324 RC6325 RC6326 RC6327 RC6328 RC6329 RC6330 RC6331 RC6332 RC6333 RC6334 RC6335 RC6336 RC6337 RC6338 RC6339 RC6340 RC6341 RC6342 RC6343 RC6344 RC6364. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
One lot of Halogen lamps have the potential to fail sooner than the expected life of the lamp which could result in damage limited to the internal electrical components inside the RetCam 3 computer module.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Clarity Medical Systems Inc
- Distributor
- Worldwide Distribution: US (nationwide) to states of: TN, MA, CA, TX, MD,FL, MS, IL, and PA; and to countries of: France, Thailand, Egypt, Mexico, Kuwait, India, Austria,Japan, Sweden, Peru, United Kingdom, Philippines,The Netherlands, China,, UAE,Israel, Germany, Italy.
- Lot
- Lot of Halogan Lamp Component is #020612C. RetCam 3, Part number 21-100500. RetCam 3 affected serial numbers: RC3066 RC3079 RC6013 RC6033 RC6034 RC6035 RC6036 RC6037 RC6038 RC6039 RC6040 RC6041 RC6043 RC6044 RC6045 RC6047 RC6049 RC6050 RC6053 RC6054 RC6055 RC6056 RC6057 RC6058 RC6059 RC6060 RC6063 RC6064 RC6066 RC6067 RC6068 RC6069 RC6070 RC6071 RC6072 RC6073 RC6074 RC6075 RC6076 RC6077 RC6078 RC6079 RC6080 RC6081 RC6082 RC6083 RC6084 RC6085 RC6086 RC6087 RC6088 RC6089 RC6090 RC6092 RC6093 RC6094 RC6095 RC6096 RC6098 RC6099 RC6101 RC6102 RC6103 RC6104 RC6105 RC6106 RC6107 RC6108 RC6109 RC6110 RC6111 RC6112 RC6113 RC6301 RC6302 RC6303 RC6304 RC6305 RC6306 RC6307 RC6309 RC6310 RC6311 RC6312 RC6313 RC6314 RC6315 RC6316 RC6317 RC6318 RC6319 RC6320 RC6321 RC6322 RC6323 RC6324 RC6325 RC6326 RC6327 RC6328 RC6329 RC6330 RC6331 RC6332 RC6333 RC6334 RC6335 RC6336 RC6337 RC6338 RC6339 RC6340 RC6341 RC6342 RC6343 RC6344 RC6364
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDADermOtic Oil (fluocinolone acetonide) 0.01% Ear Drops, 20 mL bottle, Rx only Manufactured and Distributed by:April 2, 2014
- FDAPhilips Model EPIQ 7 Ultrasound System. Intended for diagnostic ultrasound imaging and fluid flow analysis.March 26, 2014
- FDAOtto Bock Aerial and Aerial Active Wheelchair Cushions. Material numbers: SK757 476C00=SK701, 476C00=SK702,March 26, 2014
- FDAPhilips Model EPIQ 5 Ultrasound System. Intended for diagnostic ultrasound imaging and fluid flow analysis.March 26, 2014
- FDANatural-Knee® II System: POLY ULTRA TIB LT SZ00/9M POLY ULTRA TIB LT SZ0/9MM POLY ULTRA TIB LT SZ0/19M POLYMarch 26, 2014
- FDARetCam 3, RetCam Portable and RetCam Shuttle; manufactured by Clarity Medical Systems, Inc.; Pleasanton, CAMarch 19, 2014
Questions about this recall
Was Halogen Lamp component in RetCam 3 System - Clarity Medical Systems, Inc; Pleasanton, CA. General ophthalmic imaging including retinal, corneal and external imaging. Photo-documentation of pediatric ocular diseases including retinopathy of prematurity. Screening for Type-2 re-threshold retinopathy of prematurity or treatment requiring ROP. recalled?
Yes. FDA published recall Z-1265-2014 on March 26, 2014. Reported hazard: One lot of Halogen lamps have the potential to fail sooner than the expected life of the lamp which could result in damage limited to the internal electrical components inside the RetCam 3 computer module.
What should I do if I have this product?
Voluntary: Firm initiated · Letter