ARROW QuickFlash Radial Artery Catherization (wire, guide, catheter), Part Numbers: a) REF AK-04220; b) REF ASK-04220-FMH1; c) REF ASK-04220-HHC; d) REF ASK-04220-KSP; e) REF ASK-04220-PSU; f) REF ASK-04220-UCL1; g) REF ASK-04500-AH; h) REF ASK-04500-HF-S; i) REF NA-04220-S1A; j) REF NA-04220-X1A; k) REF RA-04220; l) REF RA-04220-W
Published March 27, 2024 · Recall initiated: February 12, 2024
Summary
FDA records medical device Class I recall Z-1267-2024, published March 27, 2024 and initiated February 12, 2024. Recalling firm: ARROW INTERNATIONAL Inc.; Status: Ongoing; Reported quantity: 1,160,470 devices; Lot: a) REF AK-04220, UDI/DI 20801902190943, Lot Numbers: 13F22E0436, 13F22H0360, 33F23E0375, 33F23K0032, 33F23J0443, 33F23L0862; UDI/DI 30801902121616, Lot Numbers: 13F22L0105, 13F23A0651, 33F23C0100, 33F23F0347, 33F23E0501, 33F23G0141; b) REF ASK-04220-FMH1, UDI/DI 20801902111665, Lot Numbers: 33F23B0525, 33F23H0100; c) REF ASK-04220-HHC, UDI/DI 20801902190851, Lot Numbers: 33F23F0067, 33F23K0278, 33F23G0478, 33F23H0101; d) REF ASK-04220-KSP, UDI/DI 20801902156406, Lot Numbers: 33F23E0709, 33F23H0102, 33F23H0918, 33F23J0607; e) REF ASK-04220-PSU, UDI/DI 40801902121446, Lot Numbers: 33F23F0331, 33F23G0663; f) REF ASK-04220-UCL1, UDI/DI 20801902159124, Lot Numbers: 33F23F0801, 33F23C0726, 33F23H0320; g) REF ASK-04500-AH, UDI/DI 20801902114048, Lot Numbers: 33F23E0731, 33F23H0103, 33F23H0395; h) REF ASK-04500-HF-S, UDI/DI 40801902118002, Lot Numbers: 33F23F0202, 33F23H0104, 33F23H0396; i) REF NA-04220-S1A, UDI/DI 20801902139416, Lot Numbers: 33F23J0671; j) REF NA-04220-X1A, UDI/DI 20801902136606, Lot Numbers: 33F23F0704; k) REF RA-04220, UDI/DI 20801902180678, Lot Numbers: 14F23B0041; UDI/DI 30801902193309, Lot Numbers: 14F21L0189, 14F21L0190, 14F21L0250, 14F21L0251, 14F21L0252, 14F21L0253, 14F21M0130, 14F21M0131, 14F21M0205, 14F21M0206, 14F22A0044, 14F22A0123, 14F22A0124, 14F22A0143, 14F22A0144, 14F22B0163, 14F22B0164, 14F22B0175, 14F22C0017, 14F22C0018, 14F22C0019, 14F22C0020, 14F22C0021, 14F22C0022, 14F22C0216, 14F22C0284, 14F22C0346, 14F22C0347, 14F22D0023, 14F22D0024, 14F22D0110, 14F22D0217, 14F22D0291, 14S22D0218, 14F22E0016, 14F22E0173, 14F22E0215, 14F22F0003, 14F22F0062, 14F22F0208, 14F22F0209, 14F22F0210, 14F22F0212, 14F22F0335, 14F22F0336, 14F22G0297, 14F22G0298, 14F22G0325, 14F22G0326, 14F22H0078, 14F22H0079, 14F22H0250, 14F22J0078, 14F22J0079, 14F22J0080, 14F22J0081, 14F22J0082, 14F22J0083, 14F22J0257; UDI/DI 30801902210631, Lot Numbers: 14F22K0182, 14F22K0183, 14F22K0184, 14F22K0185, 14F22K0223, 14F22M0032, 14F22M0033, 14F22M0034, 14F22M0035, 14F22M0036, 14F22M0132, 14F23B0042, 14F23B0142, 14F23B0143, 14F23B0144, 14F23B0154, 14F23D0203, 14F23D0204, 14F23D0205, 14F23D0206, 14F23D0207, 14F23E0093, 14F23E0094, 14F23E0210, 14F23E0211, 14F23F0036, 14F23F0037, 14F23F0038, 14F23F0252, 14F23F0253, 14F23F0377, 14F23F0378, 14F23F0379, 14F23G0147, 14F23G0148, 14F23G0149, 14F23G0150, 14F23G0151, 14F23G0252, 14F23H0087, 14F23H0088, 14F23H0089, 14F23H0090, 14F23H0163, 14F23H0253, 14F23H0254, 14F23H0255, 14F23J0031, 14F23J0032, 14F23J0137, 14F23J0138, 14F23J0276, 14F23J0277; l) REF RA-04220-W, UDI/DI 30801902193316, Lot Numbers: 13F22E0142, 13F22H0417, 13F22J0586, 13F22M0442, 13F23C0015, 13F23J0017; UDI/DI 30801902202988, Lot Numbers: 13F22F0124, 13F23B0363, 13F23H0010, 13F23J0013, 13F23K0001. Class I means a reasonable probability of serious adverse health consequences or death.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Teleflex is initiating this voluntary recall for the above-mentioned products due to reports received indicating a potentially defective component. The complaints received relate to resistance of the guidewire handle/chamber during use. The possible immediate health consequences of the component issue are arterial vasospasm and vessel injury arising from multiple arterial punctures with repeated attempts.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- ARROW INTERNATIONAL Inc.
- Distributor
- Worldwide distribution: US (nationwide) and OUS (foreign) to countries of: AE, AR, AT, AU, BE, BH, BR, CA, CN, CO, CY, DE, EC, EE, ES, FR, GR, HK, HR, IT, KW, LB, LU, MX, MY, NL, NZ, PH, PL, QA, RU, SA, SE, SG, SI, TH, TW, and UK.
- Lot
- a) REF AK-04220, UDI/DI 20801902190943, Lot Numbers: 13F22E0436, 13F22H0360, 33F23E0375, 33F23K0032, 33F23J0443, 33F23L0862; UDI/DI 30801902121616, Lot Numbers: 13F22L0105, 13F23A0651, 33F23C0100, 33F23F0347, 33F23E0501, 33F23G0141; b) REF ASK-04220-FMH1, UDI/DI 20801902111665, Lot Numbers: 33F23B0525, 33F23H0100; c) REF ASK-04220-HHC, UDI/DI 20801902190851, Lot Numbers: 33F23F0067, 33F23K0278, 33F23G0478, 33F23H0101; d) REF ASK-04220-KSP, UDI/DI 20801902156406, Lot Numbers: 33F23E0709, 33F23H0102, 33F23H0918, 33F23J0607; e) REF ASK-04220-PSU, UDI/DI 40801902121446, Lot Numbers: 33F23F0331, 33F23G0663; f) REF ASK-04220-UCL1, UDI/DI 20801902159124, Lot Numbers: 33F23F0801, 33F23C0726, 33F23H0320; g) REF ASK-04500-AH, UDI/DI 20801902114048, Lot Numbers: 33F23E0731, 33F23H0103, 33F23H0395; h) REF ASK-04500-HF-S, UDI/DI 40801902118002, Lot Numbers: 33F23F0202, 33F23H0104, 33F23H0396; i) REF NA-04220-S1A, UDI/DI 20801902139416, Lot Numbers: 33F23J0671; j) REF NA-04220-X1A, UDI/DI 20801902136606, Lot Numbers: 33F23F0704; k) REF RA-04220, UDI/DI 20801902180678, Lot Numbers: 14F23B0041; UDI/DI 30801902193309, Lot Numbers: 14F21L0189, 14F21L0190, 14F21L0250, 14F21L0251, 14F21L0252, 14F21L0253, 14F21M0130, 14F21M0131, 14F21M0205, 14F21M0206, 14F22A0044, 14F22A0123, 14F22A0124, 14F22A0143, 14F22A0144, 14F22B0163, 14F22B0164, 14F22B0175, 14F22C0017, 14F22C0018, 14F22C0019, 14F22C0020, 14F22C0021, 14F22C0022, 14F22C0216, 14F22C0284, 14F22C0346, 14F22C0347, 14F22D0023, 14F22D0024, 14F22D0110, 14F22D0217, 14F22D0291, 14S22D0218, 14F22E0016, 14F22E0173, 14F22E0215, 14F22F0003, 14F22F0062, 14F22F0208, 14F22F0209, 14F22F0210, 14F22F0212, 14F22F0335, 14F22F0336, 14F22G0297, 14F22G0298, 14F22G0325, 14F22G0326, 14F22H0078, 14F22H0079, 14F22H0250, 14F22J0078, 14F22J0079, 14F22J0080, 14F22J0081, 14F22J0082, 14F22J0083, 14F22J0257; UDI/DI 30801902210631, Lot Numbers: 14F22K0182, 14F22K0183, 14F22K0184, 14F22K0185, 14F22K0223, 14F22M0032, 14F22M0033, 14F22M0034, 14F22M0035, 14F22M0036, 14F22M0132, 14F23B0042, 14F23B0142, 14F23B0143, 14F23B0144, 14F23B0154, 14F23D0203, 14F23D0204, 14F23D0205, 14F23D0206, 14F23D0207, 14F23E0093, 14F23E0094, 14F23E0210, 14F23E0211, 14F23F0036, 14F23F0037, 14F23F0038, 14F23F0252, 14F23F0253, 14F23F0377, 14F23F0378, 14F23F0379, 14F23G0147, 14F23G0148, 14F23G0149, 14F23G0150, 14F23G0151, 14F23G0252, 14F23H0087, 14F23H0088, 14F23H0089, 14F23H0090, 14F23H0163, 14F23H0253, 14F23H0254, 14F23H0255, 14F23J0031, 14F23J0032, 14F23J0137, 14F23J0138, 14F23J0276, 14F23J0277; l) REF RA-04220-W, UDI/DI 30801902193316, Lot Numbers: 13F22E0142, 13F22H0417, 13F22J0586, 13F22M0442, 13F23C0015, 13F23J0017; UDI/DI 30801902202988, Lot Numbers: 13F22F0124, 13F23B0363, 13F23H0010, 13F23J0013, 13F23K0001
- Risk classification
- Class I
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAArrow Arterial Kits and Trays as follows: (1) Arrow Arterial Access Tray with 0.025" Spring-Wire GuideOctober 2, 2024
- FDAArrow UltraFlex Intra-Aortic Balloon Catheter Kit, a) REF IAB-06830-U; b) REF IAB-06840-U; c) REF IAB-06850-UJune 12, 2024
- FDAArrow Ultra 8 IAB Intra-Aortic Balloon Catheter Kit, a) REF IAB-05830-U; b) REF IAB-05840-UJune 12, 2024
- FDAArrow FiberOptix Intra-Aortic Balloon Catheter Kit: a) REF IAB-05830-LWS; b) REF IAB-05840-LWS; c) REFJune 12, 2024
- FDAWesco mint no-bake cookie bites 9oz cup UPC 750308040474March 27, 2024
- FDAsmiths medical Pneupac paraPAC plus 310 ventilator kit with PEEP and CPAP, REF P310NUSMarch 27, 2024
Questions about this recall
Was ARROW QuickFlash Radial Artery Catherization (wire, guide, catheter), Part Numbers: a) REF AK-04220; b) REF ASK-04220-FMH1; c) REF ASK-04220-HHC; d) REF ASK-04220-KSP; e) REF ASK-04220-PSU; f) REF ASK-04220-UCL1; g) REF ASK-04500-AH; h) REF ASK-04500-HF-S; i) REF NA-04220-S1A; j) REF NA-04220-X1A; k) REF RA-04220; l) REF RA-04220-W recalled?
Yes. FDA published recall Z-1267-2024 on March 27, 2024. Reported hazard: Teleflex is initiating this voluntary recall for the above-mentioned products due to reports received indicating a potentially defective component. The complaints received relate to resistance of the guidewire handle/chamber during use. The possible immediate health consequences of the component issue are arterial vasospasm and vessel injury arising from multiple arterial punctures with repeated attempts.
What should I do if I have this product?
Voluntary: Firm initiated · Letter