Back to search
FDAFDA · Food and drugs·Recall number Z-1269-2025

CircuitGuard Bacterial/Viral Filter with Elbow (Model/Catalogue number: SK200FPE / 7056)

Published March 12, 2025 · Recall initiated: January 9, 2025

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

The machine side connector of the filter HME may be occluded by plastic which prevents proper airflow. The issue may lead to ventilation bag malfunction in which the user is unable to inflate/deflate properly during manual ventilation. Potential risks to health include oxygen deprivation, leading to respiratory distress and medical complications.

What to do

Voluntary: Firm initiated · E-Mail

Recall details

Manufacturer
Kung Shin Plastics Co. Ltd.
Distributor
US Nationwide distribution in the state of IL.
Lot
UDI (01)00810050853682(10)C2N24002(11)240828(17)290827 Lot C2N24002
Risk classification
Class II
Area (FDA)
device
Country
Taiwan

Sell products like this?

Recalert watches your whole catalog for recalls like this one, automatically.

Start free — no card