Single Analyte DAU Calibrator (5 mL) Norbuprenorphine: Cutoff Calibrator; IVD - for in vitro diagnostic use only; Lin-Zhi International Sunnyvale, CA The Norbuprenorphine Drug of Abuse (DAU) cutoff Calibrator is used for qualitative and semi-qualitative calibration of the Lin-Zhi Internal, Inc. Buprenorphine (BUP) enzyme Immunoassay (REF # A53684) on the Synchron Clinical Analyzer

Published April 2, 2014 · Recall initiated: February 18, 2014

Summary

FDA records medical device Class III recall Z-1270-2014, published April 2, 2014 and initiated February 18, 2014. Recalling firm: Lin-Zhi International Inc; Status: Terminated; Reported quantity: 120 units, 60 of each lot number; Lot: Product catalog number: A68827; Lots 1308006, 1308131, Expiry: March 28, 2015. Class III means the product is not likely to cause adverse health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Customer reported that the Synchron NBUP cutoff calibrator (10 ng/mL) could not be qualified through quality control testing. The cutoff calibrator was recovering at a concentration close to the NBUP Control Level 1 (7 ng/mL).

What to do

Voluntary: Firm initiated · E-Mail

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Lin-Zhi International Inc
Distributor
US Distribution including CA.
Lot
Product catalog number: A68827; Lots 1308006, 1308131, Expiry: March 28, 2015.
Risk classification
Class III
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Single Analyte DAU Calibrator (5 mL) Norbuprenorphine: Cutoff Calibrator; IVD - for in vitro diagnostic use only; Lin-Zhi International Sunnyvale, CA The Norbuprenorphine Drug of Abuse (DAU) cutoff Calibrator is used for qualitative and semi-qualitative calibration of the Lin-Zhi Internal, Inc. Buprenorphine (BUP) enzyme Immunoassay (REF # A53684) on the Synchron Clinical Analyzer recalled?

Yes. FDA published recall Z-1270-2014 on April 2, 2014. Reported hazard: Customer reported that the Synchron NBUP cutoff calibrator (10 ng/mL) could not be qualified through quality control testing. The cutoff calibrator was recovering at a concentration close to the NBUP Control Level 1 (7 ng/mL).

What should I do if I have this product?

Voluntary: Firm initiated · E-Mail