Single Analyte DAU Calibrator (5 mL) Norbuprenorphine: Cutoff Calibrator; IVD - for in vitro diagnostic use only; Lin-Zhi International Sunnyvale, CA The Norbuprenorphine Drug of Abuse (DAU) cutoff Calibrator is used for qualitative and semi-qualitative calibration of the Lin-Zhi Internal, Inc. Buprenorphine (BUP) enzyme Immunoassay (REF # A53684) on the Synchron Clinical Analyzer
Published April 2, 2014 · Recall initiated: February 18, 2014
Summary
FDA records medical device Class III recall Z-1270-2014, published April 2, 2014 and initiated February 18, 2014. Recalling firm: Lin-Zhi International Inc; Status: Terminated; Reported quantity: 120 units, 60 of each lot number; Lot: Product catalog number: A68827; Lots 1308006, 1308131, Expiry: March 28, 2015. Class III means the product is not likely to cause adverse health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Customer reported that the Synchron NBUP cutoff calibrator (10 ng/mL) could not be qualified through quality control testing. The cutoff calibrator was recovering at a concentration close to the NBUP Control Level 1 (7 ng/mL).
What to do
Voluntary: Firm initiated · E-Mail
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Lin-Zhi International Inc
- Distributor
- US Distribution including CA.
- Lot
- Product catalog number: A68827; Lots 1308006, 1308131, Expiry: March 28, 2015.
- Risk classification
- Class III
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAAMS AST For the In vitro quantitative determination of AST in serum Quantitative determination of AST inApril 2, 2014
- FDAMEVION S250, System, Radiation Therapy, Charged-Particle, Medical used for Proton Radiation Therapy ProductApril 2, 2014
- FDAAMS URIC ACID For the In vitro quantitative determination of Uric Acid in serum Quantitative determination ofApril 2, 2014
- FDACannabinoids DAU Calibrator Control Level 1 (62.5 ng/mL) This solution is prepared from non-sterile urine andMarch 19, 2014
- FDACannabinoids DAU Calibrator Control Level 1 (37.5 ng/mL) This solution is prepared from non-sterile urine andMarch 19, 2014
- FDASingle Analyte Urine DAU Control (5 m L Norbuprenorphine, Level II control For In Vitro Diagnostic Use Only.December 4, 2013
Questions about this recall
Was Single Analyte DAU Calibrator (5 mL) Norbuprenorphine: Cutoff Calibrator; IVD - for in vitro diagnostic use only; Lin-Zhi International Sunnyvale, CA The Norbuprenorphine Drug of Abuse (DAU) cutoff Calibrator is used for qualitative and semi-qualitative calibration of the Lin-Zhi Internal, Inc. Buprenorphine (BUP) enzyme Immunoassay (REF # A53684) on the Synchron Clinical Analyzer recalled?
Yes. FDA published recall Z-1270-2014 on April 2, 2014. Reported hazard: Customer reported that the Synchron NBUP cutoff calibrator (10 ng/mL) could not be qualified through quality control testing. The cutoff calibrator was recovering at a concentration close to the NBUP Control Level 1 (7 ng/mL).
What should I do if I have this product?
Voluntary: Firm initiated · E-Mail