Back to search
FDAFDA · Food and drugs·Recall number Z-1284-2016

AXIOM Artis zee/zeego, system, x-ray, angiographic Product Usage: AXIOM Artis zee and Q/Q.zen is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients. Procedures that can be performed with the AXIOM Artis zee/zeego and Q/Q.zen family include cardiac angiography, neuro angiography, general angiography, rotational angiography, operating room angiography, multipurpose angiography and whole body radiographic/fluoroscopic procedures. AXIOM Artis zee/zeego and Q/Q.zen can also support the acquisition of position triggered imaging for spatial data synthesis.

Published April 6, 2016 · Recall initiated: February 18, 2016

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Due to a leak in the cooling system, fluids could potentially seep into the equipment cabinet. Loss of coolant can result in a cooling system failure and subsequent system failure. This leak can occur sporadically in affected systems.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Siemens Medical Solutions USA, Inc
Distributor
US Nationwide Distribution in the states of: AL, AR, AZ, CA, CO, CT, DC, FL, GA, GU, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY including Puerto Rico.
Lot
Serial Numbers: Artis Q Ceiling 109033, 109135, 109141, 109123, 109080, 109074, 109119, 109016, 109118, 109155, 109067, 109066, 109068, 109134, 109133, 109084, 109011, 109096, 109159, 109130, 109143, 109102, 109015, 109144, 109009, 109057, 109108, 109103, 109038, 109040, 109034, 109129, 109137, 109029, 109136, 109064, 109132, 109058, 109145, 109050, 109076, 109007, 109072, 109075, 109071, 109091, 109090, 109041, 109039, 109060, 109146, 109088, 109092, 109032, 109021, 109048, 109020, 109036, 109043, 109161, 109101, 109105, 109104, 109107, 109047, 109031, 109028, 109027, 109120, 109010, 109093, 109157, 109156, 109069, 109158, 109044
Risk classification
Class II
Area (FDA)
device
Country
United States

Sell products like this?

Recalert watches your whole catalog for recalls like this one, automatically.

Start free — no card