Alpha Air 6 ENT Microscope Microscope examination of the human ear, nose, and throat.
Published April 11, 2018 · Recall initiated: January 1, 2017
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The microscope may tip over if attachments are fully extended in one line and there is an attempt to move the device not by the handles mounted on the post.
What to do
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Recall details
- Manufacturer
- Seiler Instr. & Mfg. Co. Inc.
- Distributor
- Worldwide Distribution - US Distribution to the states of : CA, CO, FL, IL, MN, NV, OK, TX., and to the countries of : Canada, China, Ecuador, England, Greece, Guatemala, Jordan, Lebanon, Mexico, Romania, Russia, South Korea, Taiwan, Japan and United Arab Emirates.
- Lot
- Model Number 6 ENT
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States