Medica Wash1-Wash solution used prior to Lipase assay on the EasyRA clinical chemistry analyzer REF 10680 - Product Usage: For use in washing the probe of the Medica EasyRA Chemistry Analyzer. The wash solution in the wedge is used to prevent interference between certain tests.
Published May 15, 2019 · Recall initiated: February 6, 2019
Summary
FDA records medical device Class II recall Z-1294-2019, published May 15, 2019 and initiated February 6, 2019. Recalling firm: Medica Corporation; Status: Terminated; Reported quantity: 87 kits; Lot: Lot #s 18066, 18133, 18235, 18317. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Wash1 solution is ineffective in cleaning dispensing probe of TRIG and LDL reagents that interfere with the Lipase (LIP) assay on the EasyRA clinical chemistry analyzer.
What to do
Voluntary: Firm initiated · Telephone
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Medica Corporation
- Distributor
- Worldwide distribution - US Nationwide and countries of IRAQ, FRANCE, INDIA, GEORGIA, GEORGIA, FRANCE GUATEMALA, KENYA, IRAN, ESTONIA.
- Lot
- Lot #s 18066, 18133, 18235, 18317
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAVITROS® 3600 Immunoassay System-Software V3.3.2 & below Product Code: 6802783 For use in the in vitroMay 15, 2019
- FDACios Alpha with software version VA30-mobile C-arm x-ray Material # 11105200 Product Usage: The Cios Alpha isMay 15, 2019
- FDAEl Guapo Pasilla-Ancho Chile Pods 2oz, 7.5oz, 11ozMay 15, 2019
- FDAT7 Driver Cannulated AO (Part Number 110018531) Product Usage: The Biomet Headless Compression Screws andMay 15, 2019
- FDAFresh cut watermelon, fresh cut honeydew melon, fresh cut cantaloupe and fresh cut mixed fruit containing oneMay 15, 2019
- FDAConMed Anchor Tissue Retrieval System 10 MM, 235 ML, (5/BX) Catalog Number: TRS100SB2 The Anchor TissueMay 15, 2019
Questions about this recall
Was Medica Wash1-Wash solution used prior to Lipase assay on the EasyRA clinical chemistry analyzer REF 10680 - Product Usage: For use in washing the probe of the Medica EasyRA Chemistry Analyzer. The wash solution in the wedge is used to prevent interference between certain tests. recalled?
Yes. FDA published recall Z-1294-2019 on May 15, 2019. Reported hazard: Wash1 solution is ineffective in cleaning dispensing probe of TRIG and LDL reagents that interfere with the Lipase (LIP) assay on the EasyRA clinical chemistry analyzer.
What should I do if I have this product?
Voluntary: Firm initiated · Telephone