Back to search
FDAFDA · Food and drugs·Recall number Z-1311-2026

Brand Name: Molift Product Name: Molift 4-point sling bars (Accessory to mobile hoists) Model/Catalog Number: 1430116,1430117,1430115 Product Description: The mobile hoist is an assistive device intended for alleviation of, or compensation for, a functional, impairment due to an injury or disability. The device is designed for an individual lacking the ability to stand up and transfer themselves to / from a bed, the floor, a wheelchair, a chair, a toilet or similar due to reduced mobility or physical strength.

Published February 11, 2026 · Recall initiated: November 17, 2025

Summary

FDA records medical device Class II recall Z-1311-2026, published February 11, 2026 and initiated November 17, 2025. Recalling firm: ETAC A/S; Status: Ongoing; Reported quantity: 25 units; Lot: (1) Model/Catalog Number: 1430116; UDI-DI: TBD; (2) Model/Catalog Number: 1430117; UDI-DI: TBD; (3) Model/Catalog Number: 1430115; UDI-DI: TBD. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

A customer experienced that during assembly and load testing of the device, the bolt for mounting the lifting bar broke. Upon further investigation of this customer complaint, it became clear that the defect occurs as a result of deformation of the materials involved between the lifting bar and the arm of the device. Such deformation can lead to limited mobility in the joint, which prevents proper functionality of the equipment.

What to do

Voluntary: Firm initiated · Telephone

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
ETAC A/S
Distributor
Worldwide - US Nationwide distribution in the state of PA and the countries of Australia, Czech Republic, Finland, France, Germany, Israel, Italy, Japan, New Zealand, Norway, Poland, Qatar Spain, Sweden, Switzerland, UK.
Lot
(1) Model/Catalog Number: 1430116; UDI-DI: TBD; (2) Model/Catalog Number: 1430117; UDI-DI: TBD; (3) Model/Catalog Number: 1430115; UDI-DI: TBD;
Risk classification
Class II
Area (FDA)
device
Country
Denmark

Related recalls

Questions about this recall

Was Brand Name: Molift Product Name: Molift 4-point sling bars (Accessory to mobile hoists) Model/Catalog Number: 1430116,1430117,1430115 Product Description: The mobile hoist is an assistive device intended for alleviation of, or compensation for, a functional, impairment due to an injury or disability. The device is designed for an individual lacking the ability to stand up and transfer themselves to / from a bed, the floor, a wheelchair, a chair, a toilet or similar due to reduced mobility or physical strength. recalled?

Yes. FDA published recall Z-1311-2026 on February 11, 2026. Reported hazard: A customer experienced that during assembly and load testing of the device, the bolt for mounting the lifting bar broke. Upon further investigation of this customer complaint, it became clear that the defect occurs as a result of deformation of the materials involved between the lifting bar and the arm of the device. Such deformation can lead to limited mobility in the joint, which prevents proper functionality of the equipment.

What should I do if I have this product?

Voluntary: Firm initiated · Telephone