AOA Distalizer Lock Nut Screw, found in products listed as: Distal Jet, Spring Jet, Mesial Jet, and Uprighter Jet. Removable functional orthodontic appliances that are to modify the growth of the jaws in a prescribed growth pattern.
Published April 9, 2014 · Recall initiated: February 24, 2014
Summary
FDA records medical device Class II recall Z-1314-2014, published April 9, 2014 and initiated February 24, 2014. Recalling firm: Allesee Orthodontic Appliances; Status: Terminated; Reported quantity: 411; Lot: All Devices/Codes Manufactured from August 28, 2013 to November 6, 2013. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The Lock Nut component of the Jet appliances may be missing the screw collar, which serves as a mechanical stop to prevent over-tightening in appliance assemby. Without this collar, an over-tightened lock nut may stop the piston assembly from moving freely, subsequently preventing intended movement of the teeth.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Allesee Orthodontic Appliances
- Distributor
- Worldwide Distribution: US (nationwide) and the country of : Canada.
- Lot
- All Devices/Codes Manufactured from August 28, 2013 to November 6, 2013
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAInfantino Squeeze & Teethe Monkey, "Go Gaga Squeeze & Teether Coco the Monkey". Non-Fluid Filled TeethingApril 9, 2014
- FDAOmni Micro-electrode/reference electrode for cobas b221 analyzer, Model/Catalog/Part Number: 03111873180 as aApril 9, 2014
- FDAPinnacle3 Radiation Therapy Planning (RTP) System, software version 9.0.April 9, 2014
- FDAIMMULITE 2000; IMMULITE 2000XPi; VersaCell The IMMULITE 2000 systems are continuous random-access instrumentsApril 9, 2014
- FDAZoll Medical X Series, PropaqMD, PropaqM Defibrillator/Pacemaker/Monitor With System Software VersionApril 9, 2014
- FDACitation TMZF HA 132 degrees Neck Angle, V40 Hip Stem - Right Size #3: catalog # 6265-5113 , Left Size #4:April 9, 2014
Questions about this recall
Was AOA Distalizer Lock Nut Screw, found in products listed as: Distal Jet, Spring Jet, Mesial Jet, and Uprighter Jet. Removable functional orthodontic appliances that are to modify the growth of the jaws in a prescribed growth pattern. recalled?
Yes. FDA published recall Z-1314-2014 on April 9, 2014. Reported hazard: The Lock Nut component of the Jet appliances may be missing the screw collar, which serves as a mechanical stop to prevent over-tightening in appliance assemby. Without this collar, an over-tightened lock nut may stop the piston assembly from moving freely, subsequently preventing intended movement of the teeth.
What should I do if I have this product?
Voluntary: Firm initiated · Letter