FDAFDA · Food and drugs·Recall number Z-1314-2024
Abbott Point of Care Control Control i-STAT¿ Level 1 10 X 1.7 mL
Published March 27, 2024 · Recall initiated: January 9, 2024
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Facility Critical Temperature Excursions on 08-JAN 2024 in Walk-In refrigerator due to mechanical failures, causing temperatures to fall below the validated range of 35.6 - 46.4 Degrees Fahrenheit (2 - 8 Degrees Celsius) for up to 60 minutes which could increase the probability of an incorrect test result or a delayed test result.
What to do
Voluntary: Firm initiated · Telephone
Recall details
- Manufacturer
- Mckesson Medical-Surgical Inc. Corporate Office
- Distributor
- US Nationwide distribution in the states of AZ, PA and TX.
- Lot
- Catalogue Number: 06F1201; UDI/DI: 54749000463; Lot No: 101164
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States