Back to search
FDAFDA · Food and drugs·Recall number Z-1314-2024

Abbott Point of Care Control Control i-STAT¿ Level 1 10 X 1.7 mL

Published March 27, 2024 · Recall initiated: January 9, 2024

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Facility Critical Temperature Excursions on 08-JAN 2024 in Walk-In refrigerator due to mechanical failures, causing temperatures to fall below the validated range of 35.6 - 46.4 Degrees Fahrenheit (2 - 8 Degrees Celsius) for up to 60 minutes which could increase the probability of an incorrect test result or a delayed test result.

What to do

Voluntary: Firm initiated · Telephone

Recall details

Manufacturer
Mckesson Medical-Surgical Inc. Corporate Office
Distributor
US Nationwide distribution in the states of AZ, PA and TX.
Lot
Catalogue Number: 06F1201; UDI/DI: 54749000463; Lot No: 101164
Risk classification
Class II
Area (FDA)
device
Country
United States