Back to search
FDAFDA · Food and drugs·Recall number Z-1328-2025

Stryker CinchLock Flex Knotless Anchor with Inserter and Suture, REF CAT02643, Sterile EO, Rx Only MR Safe CE 0197

Published March 19, 2025 · Recall initiated: January 29, 2025

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Increase in complaints concerning "pullwire" breakage that may result in the broken pullwire remaining in the anchor implant body after deployment.

What to do

Voluntary: Firm initiated

Recall details

Manufacturer
Stryker Corporation
Distributor
Worldwide - US Nationwide distribution including in the states of Alabama, Alaska, Arizona, Arkansas, California, Colorado, Connecticut, Delaware, Florida, Georgia, Hawaii, Idaho, Illinois, Indiana, Iowa, Kansas, Kentucky, Louisiana, Maine, Maryland, Michigan, Minnesota, Mississippi, Missouri, Montana, Nebraska, Nevada, New Hampshire, New Jersey, New York, North Carolina, North Dakota, Ohio, Oklahoma, Oregon, Pennsylvania, Puerto Rico, South Carolina, South Dakota, Tennessee, Texas, Utah, Virginia, Washington, West Virginia, Wisconsin, Wyoming and the countries of Argentina, Colombia, Netherlands, Spain, Mexico, Australia, United Kingdom, Brazil.
Lot
UDI:07613327123661/ Lot #s: 23342AE2 24190AE2 23345AE2 24200AE2 24004AE2 24201AE2 24008AE2 24204AE2 24009AE2 24205AE2 24010AE2 24206AE2 24032AE2 24214AE2 24033AE2 24215AE2 24036AE2 24221AE2 24050AE2 24222AE2 24061AE2 24226AE2 24064AE2 24233AE2 24065AE2 24235AE2 24066AE2 24240AE2 24087AE2 24248AE2 24093AE2 24249AE2 24094AE2 24250AE2 24127AE2 24256AE2 24128AE2 24257AE2 24135AE2 24258AE2 24136AE2 24260AE2 24137AE2 24276AE2 24143AE2 24277AE2 24144AE2 24278AE2 24159AE2 24283AE2 24160AE2 24284AE2 24162AE2 24285AE2 24169AE2 24292AE2 24170AE2 24306AE2 24184AE2 24309AE2 24185AE2 24311AE2 24186AE2 24312AE2