Back to search
FDAFDA · Food and drugs·Recall number Z-1354-2015

Syngo.plaza. A Picture Archiving and Communication System (PACS) intended to display, process, read, report, communicate, distribute, store and archive digital medical images. It supports the physician in diagnosis and treatment planning.

Published April 8, 2015 · Recall initiated: February 18, 2015

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Possibly incomplete archived studies during pre-fetch. In a server farm setup, when pre-fetch/retrieve operation is performed for partially archived studies, the series that have not yet been archived, will remain unarchived.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Siemens Medical Solutions USA, Inc
Distributor
US Distribution: MA, WI, FL
Lot
Model Numbers of device: 10592457, 10863172, serial numbers: 100177 100033 100181 100029
Risk classification
Class II
Area (FDA)
device
Country
United States