Back to search
FDAFDA · Food and drugs·Recall number Z-1360-2023

Titan Pump Assembly, Catalog Number 5177701400; inflatable penile Prosthesis

Published April 19, 2023 · Recall initiated: March 1, 2023

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

A decrease in wall thickness has the potential for a premature pump failure, compared to a pump with standard wall thickness, based on the number of interactions with the pump to inflate and deflate the penile prosthesis.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Coloplast Manufacturing US, LLC
Distributor
Worldwide distribution - US Nationwide and the countries of Turkey, Italy, Germany, Belgium, France, United Kingdom, Spain, Portugal.
Lot
UDI/DI 05708932319621, Lot Numbers: 8904177, 8849617
Risk classification
Class II
Area (FDA)
device
Country
United States