CareFusion Alaris PCA Module, Model 8120
Published March 11, 2020 · Recall initiated: February 4, 2020
Summary
FDA records medical device Class I recall Z-1362-2020, published March 11, 2020 and initiated February 4, 2020. Recalling firm: CareFusion 303, Inc.; Status: Ongoing; Reported quantity: 83015; Lot: Software versions 12.1.0, and 9.33 and prior. Class I means a reasonable probability of serious adverse health consequences or death.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
1)Error 255-XX-XXX results in inability to edit settings 2)Delay options programming may result in no KVO rate/therapy interruption 3)Software errors results in no low battery alarm/infusion stopping 4)Medium priority KVO/End of Infusion alarms may result in unrecognized infusion completion 5)Custom concentration data entry errors results in concentrations being lower/higher than medication orders
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- CareFusion 303, Inc.
- Distributor
- Worldwide Distribution: (U.S. Nationwide) to states of: CO, SC, NY, CA, LA, TX, AK, ID, VA, DE, FL, PA, KY, KS, TN, WA, MS, IL, GA, AR, NC, NV, NE, OR, RI, WI, HI, GU, PR, MA, IA, CT, MI, AZ, ME, MT, IN, OH, WV, WY, MN, AL, NH, OK, DC, NJ, MO, SD, UT, NM, MD, VT, MP, ND, AS; and countries of: Canada, Australia, New Zealand, the Middle East, South Africa, Taiwan, Southeast Asia and the European Union.
- Lot
- Software versions 12.1.0, and 9.33 and prior
- Risk classification
- Class I
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDABD Pyxis Anesthesia Station 4000 REF: 338 Medication cabinetJune 10, 2026
- FDABD Pyxis MedBank Tower REF: 1145-00 Medication cabinetJune 10, 2026
- FDABD Pyxis Mini Main REF: 349 Medication cabinetJune 10, 2026
- FDABD Pyxis MedStation ES REF: 323 Medication cabinetJune 10, 2026
- FDACareFusion Alaris System, Alaris PC unit models 8000 and 8015 (UDI: 10885403801518)March 11, 2020
- FDADeRoyal Sterile Custom Kits containing Cardinal Health Gowns. Sold as the following: ASC Thyroid PackMarch 11, 2020
Questions about this recall
Was CareFusion Alaris PCA Module, Model 8120 recalled?
Yes. FDA published recall Z-1362-2020 on March 11, 2020. Reported hazard: 1)Error 255-XX-XXX results in inability to edit settings 2)Delay options programming may result in no KVO rate/therapy interruption 3)Software errors results in no low battery alarm/infusion stopping 4)Medium priority KVO/End of Infusion alarms may result in unrecognized infusion completion 5)Custom concentration data entry errors results in concentrations being lower/higher than medication orders
What should I do if I have this product?
Voluntary: Firm initiated · Letter