Boston Scientific TSX Transseptal Needle, Model Number M004TSX60o; Used to create the primary puncture in the interatrial septum when passing an intruder and/or catheter through the septum from the right side of the heart to the left side.
Published April 20, 2016 · Recall initiated: March 15, 2016
Summary
FDA records medical device Class II recall Z-1375-2016, published April 20, 2016 and initiated March 15, 2016. Recalling firm: Merit Medical Systems, Inc.; Status: Terminated; Reported quantity: 25; Lot: Lot Q823423, Exp 7/31/2018. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Merit Medical Systems, Inc. is voluntarily conducting a recall of one lot of the TSX" Transseptal Needle due to a discrepancy between the labeling and the actual configuration of the device in the packaging. The curve dimensional values on the labels of this lot incorrectly indicate a 50® curve, rather than the correct 86® curve
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Merit Medical Systems, Inc.
- Distributor
- Distributed to one customer in the US.
- Lot
- Lot Q823423, Exp 7/31/2018
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAScout MD, REF MDS05A-01, MDS05C-01, MDS10C-01 Contents: (1) Scout MD Delivery System, Standard for use withOctober 21, 2026
- FDAEmbosphere 500-700 micrometer 2 mL Syringe Sterile Steam, REF: S420GH/EUA, S620GH/EUA; Embosphere 700-900October 7, 2026
- FDASplashwire Hydrophilic Guide Wire: REF: MSWSTDA35150 UDI-DI codes: 00884450529645 00884450529669June 24, 2026
- FDAPhilips Bridge Prep Kit REF: K12-09098C Sterile EO, Rx ONLYMay 6, 2026
- FDARotaflow Centrifugal Pump Custom tubing packs for the oxygenatorApril 20, 2016
- FDASurgiCounter scanner, part of the SurgiCount Safety-Sponge System. Model numbers 0694-097-000 andApril 20, 2016
Questions about this recall
Was Boston Scientific TSX Transseptal Needle, Model Number M004TSX60o; Used to create the primary puncture in the interatrial septum when passing an intruder and/or catheter through the septum from the right side of the heart to the left side. recalled?
Yes. FDA published recall Z-1375-2016 on April 20, 2016. Reported hazard: Merit Medical Systems, Inc. is voluntarily conducting a recall of one lot of the TSX" Transseptal Needle due to a discrepancy between the labeling and the actual configuration of the device in the packaging. The curve dimensional values on the labels of this lot incorrectly indicate a 50® curve, rather than the correct 86® curve
What should I do if I have this product?
Voluntary: Firm initiated · Letter