***REF S-3D*** 3mm Diamond Ball. Used for cutting and shaping bone.
Published June 5, 2013 · Recall initiated: March 4, 2011
Summary
FDA records medical device Class II recall Z-1377-2013, published June 5, 2013 and initiated March 4, 2011. Recalling firm: The Anspach Effort, Inc.; Status: Terminated; Reported quantity: 1,827 burrs; Lot: Lots numbers: D433040800, D403040236, D383039472. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Due to three batches of product may have diamond grit size larger than intended applied to the head of the cutting burr.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- The Anspach Effort, Inc.
- Distributor
- Worldwide Distribution-USA (nationwide) including DC and the states of OH, TX, IL, CA, GA, KY, FL, WI, NC, VA, WI, MN, MD, KS, OK, PA, OR ,HI, AZ, WA, TN, AR, IA, MI, CO, LA, NJ, IN, WV, MO, AL, NY, VT, and SC and the countries of France, Canada, Great Britain, Mexico, Australia, Sweden and Netherlands.
- Lot
- Lots numbers: D433040800, D403040236, D383039472.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAElectric Handpiece EM Hand Control Product Usage: Electric systemMay 10, 2017
- FDA14cm QD Angle Attachment Product Usage: Pneumatic systemMay 10, 2017
- FDAKeyless Driver Product Usage: Pneumatic systemMay 10, 2017
- FDA6.8cm Controlled Depth Attachment Product Usage: Pneumatic systemMay 10, 2017
- FDANEXGEN Complete Knee Solution Monoblock Tibial Drill with stop 10.7 mm. Used to prepare cylindrical holes inJune 5, 2013
- FDAFemoral component, CR, cemented, #1, left 2103-1310 Femoral component, CR, cemented, #2, left 2103-1320June 5, 2013
Questions about this recall
Was ***REF S-3D*** 3mm Diamond Ball. Used for cutting and shaping bone. recalled?
Yes. FDA published recall Z-1377-2013 on June 5, 2013. Reported hazard: Due to three batches of product may have diamond grit size larger than intended applied to the head of the cutting burr.
What should I do if I have this product?
Voluntary: Firm initiated · Letter