Check-Flo Performer Introducer Set, Individual product is packaged in a Tyvek-film sterilizable outer pouch. Introducers are intended for introduction of balloons, closed and non-tapered end catheters or other diagnostic and interventional devices.
Published March 8, 2017 · Recall initiated: February 1, 2017
Summary
FDA records medical device Class II recall Z-1388-2017, published March 8, 2017 and initiated February 1, 2017. Recalling firm: Cook Inc.; Status: Terminated; Reported quantity: 854; Lot: Catalog Identifier:RCF-4.0(5.0,6.0,7.0,8.0)- / -BV Codes consist of Lot #, RPN #, Manufacture Date and Expiration Date: 4860671 RCF-4.0-35-J-BV 2014/03 3/1/2017; 5018663 RCF-4.0-35-J-BV 6/1/2014 6/1/2017; 5152571 RCF-4.0-35-J-BV 8/5/2014 8/5/2017; 5219529 RCF-4.0-35-J-BV 9/4/2014 9/4/2017; 5475523 RCF-4.0-35-J-BV 12/9/2014 12/9/2017; 6125807 RCF-4.0-35-J-BV 9/8/2015 9/8/2018; NS5210596 RCF-4.0-35-J-BV 8/27/2014 8/27/2017; 4790887 RCF-5.0-35-J-BV 2014/02 2/1/2017; 4879368 RCF-5.0-35-J-BV 2014/04 4/1/2017; 4995370 RCF-5.0-35-J-BV 5/1/2014 5/1/2017; 5018661 RCF-5.0-35-J-BV 6/1/2014 6/1/2017; 5105572 RCF-5.0-35-J-BV 7/31/2014 7/31/2017; 5191953 RCF-5.0-35-J-BV 8/14/2014 8/14/2017; 5230090 RCF-5.0-35-J-BV 9/4/2014 9/4/2017; 5335405 RCF-5.0-35-J-BV 10/22/2014 10/22/2017; 5391106 RCF-5.0-35-J-BV 11/18/2014 11/18/2017; 5550365 RCF-5.0-35-J-BV 2/3/2015 2/3/2018; 5632266 RCF-5.0-35-J-BV 3/3/2015 3/3/2018; 5825763 RCF-5.0-35-J-BV 5/8/2015 5/8/2018; NS6041240 RCF-5.0-35-J-BV 8/6/2015 8/6/2018; 4800069 RCF-6.0-35-J-BV 2014/02 2/1/2017; 4938694 RCF-6.0-35-J-BV 2014/05 5/1/2017; 4995361 RCF-6.0-35-J-BV 6/1/2014 6/1/2017; 5026341 RCF-6.0-35-J-BV 6/1/2014 6/1/2017; 5105539 RCF-6.0-35-J-BV 7/18/2014 7/18/2017; 5115956 RCF-6.0-35-J-BV 7/23/2014 7/23/2017; 5191946 RCF-6.0-35-J-BV 8/15/2014 8/15/2017; 5204970 RCF-6.0-35-J-BV 8/28/2014 8/28/2017; 5237288 RCF-6.0-35-J-BV 9/12/2014 9/12/2017; 5306462 RCF-6.0-35-J-BV 10/3/2014 10/3/2017; 5335304 RCF-6.0-35-J-BV 10/20/2014 10/20/2017; 5589594 RCF-6.0-35-J-BV 2/23/2015 2/23/2018; 5621222 RCF-6.0-35-J-BV 2/24/2015 2/24/2018; 5781933 RCF-6.0-35-J-BV 4/30/2015 4/30/2018; 5839252 RCF-6.0-35-J-BV 5/11/2015 5/11/2018; 5933749 RCF-6.0-35-J-BV 7/10/2015 7/10/2018; 5998390 RCF-6.0-35-J-BV 7/28/2015 7/28/2018; 6145159 RCF-6.0-35-J-BV 9/11/2015 9/11/2018; NS6043629 RCF-6.0-35-J-BV 8/9/2015 8/9/2018; NS6332956 RCF-6.0-35-J-BV 11/17/2015 11/17/2018; 4790871 RCF-7.0-35-J-BV 2014/03 3/1/2017; 5018638 RCF-7.0-35-J-BV 6/1/2014 6/1/2017; 5459181 RCF-7.0-35-J-BV 12/11/2014 12/11/2017; 5600122 RCF-7.0-35-J-BV 2/6/2015 2/6/2018; 5765286 RCF-7.0-35-J-BV 4/20/2015 4/20/2018; 5905965 RCF-7.0-35-J-BV 6/15/2015 6/15/2018; 4729143 RCF-8.0-38-J-BV 2014/01 1/1/2017; 4893493 RCF-8.0-38-J-BV 2014/04 4/1/2017; 5018702 RCF-8.0-38-J-BV 6/1/2014 6/1/2017; 5214240 RCF-8.0-38-J-BV 8/26/2014 8/26/2017; 5429421 RCF-8.0-38-J-BV 11/18/2014 11/18/2017; 5486519 RCF-8.0-38-J-BV 1/6/2015 1/6/2018; 5637340 RCF-8.0-38-J-BV 3/2/2015 3/2/2018; 6073269 RCF-8.0-38-J-BV 8/13/2015 8/13/2018. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Devices using a specific hemostatic valve design, referred to as a blue valve or polyisoprene valve, have an increase in reports of blood loss associated with the devices which could result in delay in procedure and blood loss from patients.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Cook Inc.
- Distributor
- Worldwide Distribution - US including AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV and Internationally to Canada, Armenia, Austria, Belgium, Bulgaria, Cyprus, Czech Republic, Denmark, Spain, Finland, France, Germany, Greece, Hungary, Ireland, Italy, Kenya, Kuwait, Lithuania, Mauritius, Netherlands, Norway, Oman, Poland, Reunion, Romania, Saudi Arabia, Slovakia, Slovenia, South Africa, Sweden, Switzerland, Turkey, United Arab Emirates, and United Kingdom VA/DOD: Yes
- Lot
- Catalog Identifier:RCF-4.0(5.0,6.0,7.0,8.0)- / -BV Codes consist of Lot #, RPN #, Manufacture Date and Expiration Date: 4860671 RCF-4.0-35-J-BV 2014/03 3/1/2017; 5018663 RCF-4.0-35-J-BV 6/1/2014 6/1/2017; 5152571 RCF-4.0-35-J-BV 8/5/2014 8/5/2017; 5219529 RCF-4.0-35-J-BV 9/4/2014 9/4/2017; 5475523 RCF-4.0-35-J-BV 12/9/2014 12/9/2017; 6125807 RCF-4.0-35-J-BV 9/8/2015 9/8/2018; NS5210596 RCF-4.0-35-J-BV 8/27/2014 8/27/2017; 4790887 RCF-5.0-35-J-BV 2014/02 2/1/2017; 4879368 RCF-5.0-35-J-BV 2014/04 4/1/2017; 4995370 RCF-5.0-35-J-BV 5/1/2014 5/1/2017; 5018661 RCF-5.0-35-J-BV 6/1/2014 6/1/2017; 5105572 RCF-5.0-35-J-BV 7/31/2014 7/31/2017; 5191953 RCF-5.0-35-J-BV 8/14/2014 8/14/2017; 5230090 RCF-5.0-35-J-BV 9/4/2014 9/4/2017; 5335405 RCF-5.0-35-J-BV 10/22/2014 10/22/2017; 5391106 RCF-5.0-35-J-BV 11/18/2014 11/18/2017; 5550365 RCF-5.0-35-J-BV 2/3/2015 2/3/2018; 5632266 RCF-5.0-35-J-BV 3/3/2015 3/3/2018; 5825763 RCF-5.0-35-J-BV 5/8/2015 5/8/2018; NS6041240 RCF-5.0-35-J-BV 8/6/2015 8/6/2018; 4800069 RCF-6.0-35-J-BV 2014/02 2/1/2017; 4938694 RCF-6.0-35-J-BV 2014/05 5/1/2017; 4995361 RCF-6.0-35-J-BV 6/1/2014 6/1/2017; 5026341 RCF-6.0-35-J-BV 6/1/2014 6/1/2017; 5105539 RCF-6.0-35-J-BV 7/18/2014 7/18/2017; 5115956 RCF-6.0-35-J-BV 7/23/2014 7/23/2017; 5191946 RCF-6.0-35-J-BV 8/15/2014 8/15/2017; 5204970 RCF-6.0-35-J-BV 8/28/2014 8/28/2017; 5237288 RCF-6.0-35-J-BV 9/12/2014 9/12/2017; 5306462 RCF-6.0-35-J-BV 10/3/2014 10/3/2017; 5335304 RCF-6.0-35-J-BV 10/20/2014 10/20/2017; 5589594 RCF-6.0-35-J-BV 2/23/2015 2/23/2018; 5621222 RCF-6.0-35-J-BV 2/24/2015 2/24/2018; 5781933 RCF-6.0-35-J-BV 4/30/2015 4/30/2018; 5839252 RCF-6.0-35-J-BV 5/11/2015 5/11/2018; 5933749 RCF-6.0-35-J-BV 7/10/2015 7/10/2018; 5998390 RCF-6.0-35-J-BV 7/28/2015 7/28/2018; 6145159 RCF-6.0-35-J-BV 9/11/2015 9/11/2018; NS6043629 RCF-6.0-35-J-BV 8/9/2015 8/9/2018; NS6332956 RCF-6.0-35-J-BV 11/17/2015 11/17/2018; 4790871 RCF-7.0-35-J-BV 2014/03 3/1/2017; 5018638 RCF-7.0-35-J-BV 6/1/2014 6/1/2017; 5459181 RCF-7.0-35-J-BV 12/11/2014 12/11/2017; 5600122 RCF-7.0-35-J-BV 2/6/2015 2/6/2018; 5765286 RCF-7.0-35-J-BV 4/20/2015 4/20/2018; 5905965 RCF-7.0-35-J-BV 6/15/2015 6/15/2018; 4729143 RCF-8.0-38-J-BV 2014/01 1/1/2017; 4893493 RCF-8.0-38-J-BV 2014/04 4/1/2017; 5018702 RCF-8.0-38-J-BV 6/1/2014 6/1/2017; 5214240 RCF-8.0-38-J-BV 8/26/2014 8/26/2017; 5429421 RCF-8.0-38-J-BV 11/18/2014 11/18/2017; 5486519 RCF-8.0-38-J-BV 1/6/2015 1/6/2018; 5637340 RCF-8.0-38-J-BV 3/2/2015 3/2/2018; 6073269 RCF-8.0-38-J-BV 8/13/2015 8/13/2018.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAFlexor Check-Flo Introducer - Raabe Modification, Introducer Set, RPN KCFW-5.0-38-90-RB-RAABE, GPN G13069 TheNovember 10, 2021
- FDAFlexor Check-Flo Introducer, RPN KCFW-8.0-35-55-RB-HFANL0-HC, Order Number G56233 Flexor Introducers andDecember 30, 2020
- FDAFlexor Check-Flo Introducer, RPN KCFW-6.0-35-55-RB-HFANL0-HC, Order Number G56231 Flexor Introducers andDecember 30, 2020
- FDAFlexor Check-Flo Introducer, RPN KCFW-6.0-35-55-RB-HFANL1-HC, Order Number G56236 Flexor Introducers andDecember 30, 2020
- FDACheck-Flo Performer Introducer, Individual product is packaged in a Tyvek-film sterilizable outer pouch.March 8, 2017
- FDACheck-Flo Hemostasis Assembly, Individual product is packaged in a Tyvek-film sterilizable outer pouch.March 8, 2017
Questions about this recall
Was Check-Flo Performer Introducer Set, Individual product is packaged in a Tyvek-film sterilizable outer pouch. Introducers are intended for introduction of balloons, closed and non-tapered end catheters or other diagnostic and interventional devices. recalled?
Yes. FDA published recall Z-1388-2017 on March 8, 2017. Reported hazard: Devices using a specific hemostatic valve design, referred to as a blue valve or polyisoprene valve, have an increase in reports of blood loss associated with the devices which could result in delay in procedure and blood loss from patients.
What should I do if I have this product?
Voluntary: Firm initiated · Letter