Stryker Orthobiologics Aliquot Delivery System Manual Surgical Instrument - Aliquot Catheter, 6", Part Number 2110-0503 Product Usage: Facilitate placement of bone cement in weakened or diminished bone

Published June 5, 2013 · Recall initiated: November 10, 2012

Summary

FDA records medical device Class II recall Z-1399-2013, published June 5, 2013 and initiated November 10, 2012. Recalling firm: Orthovita, Inc., dBA Stryker Orthobiologics.; Status: Terminated; Reported quantity: 69; Lot: Manufacturer Part Number 2110-0503, Lot Numbers E908003R and E908002. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

There have been a few reported cases of the catheter breaking when removing the Aliquot catheter from the needle during vertebroplasty procedures. This has been attributed to bending of the catheter during the retraction from the needle. It was determined to be use errors. The correction is limited to a revision of the Directions for Use.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Orthovita, Inc., dBA Stryker Orthobiologics.
Distributor
USA Nationwide Distribution in the states of AZ, CA, CO, CT, DE, FL, GA, HI, IL, IN, KS, KY, LA, MD, ME, MI, MO, NC, ND, NH, NJ, NY, OH, OK, PA, RI, SC, TN, TX, UT, VA, WA, and WI.
Lot
Manufacturer Part Number 2110-0503, Lot Numbers E908003R and E908002
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Stryker Orthobiologics Aliquot Delivery System Manual Surgical Instrument - Aliquot Catheter, 6", Part Number 2110-0503 Product Usage: Facilitate placement of bone cement in weakened or diminished bone recalled?

Yes. FDA published recall Z-1399-2013 on June 5, 2013. Reported hazard: There have been a few reported cases of the catheter breaking when removing the Aliquot catheter from the needle during vertebroplasty procedures. This has been attributed to bending of the catheter during the retraction from the needle. It was determined to be use errors. The correction is limited to a revision of the Directions for Use.

What should I do if I have this product?