Stryker Orthobiologics Aliquot Delivery System Manual Surgical Instrument - Aliquot Catheter, 6", Part Number 2110-0503 Product Usage: Facilitate placement of bone cement in weakened or diminished bone
Published June 5, 2013 · Recall initiated: November 10, 2012
Summary
FDA records medical device Class II recall Z-1399-2013, published June 5, 2013 and initiated November 10, 2012. Recalling firm: Orthovita, Inc., dBA Stryker Orthobiologics.; Status: Terminated; Reported quantity: 69; Lot: Manufacturer Part Number 2110-0503, Lot Numbers E908003R and E908002. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
There have been a few reported cases of the catheter breaking when removing the Aliquot catheter from the needle during vertebroplasty procedures. This has been attributed to bending of the catheter during the retraction from the needle. It was determined to be use errors. The correction is limited to a revision of the Directions for Use.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Orthovita, Inc., dBA Stryker Orthobiologics.
- Distributor
- USA Nationwide Distribution in the states of AZ, CA, CO, CT, DE, FL, GA, HI, IL, IN, KS, KY, LA, MD, ME, MI, MO, NC, ND, NH, NJ, NY, OH, OK, PA, RI, SC, TN, TX, UT, VA, WA, and WI.
- Lot
- Manufacturer Part Number 2110-0503, Lot Numbers E908003R and E908002
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAVitagel; indicated in surgical procedures (other than in neurosurgical and ophthalmic) as an adjunct toApril 8, 2015
- FDAStryker Orthobiologics Aliquot Delivery System Manual Surgical Instrument - Aliquot Plunger Delivery System,June 5, 2013
- FDAStryker Orthobiologics Aliquot Delivery System Manual Surgical Instrument - Aliquot Reamer, 6", Part NumberJune 5, 2013
- FDAStryker Orthobiologics Aliquot Delivery System Manual Surgical Instrument - Aliquot Plunger, 6", Part NumberJune 5, 2013
- FDAStryker Orthobiologics Aliquot Delivery System Manual Surgical Instrument - Aliquot Beveled Needle, 6", PartJune 5, 2013
- FDASynchron LX System(s) and UniCel DxC 600/800 System(s) Lactate Reagent, Part Number: A95550 Product Usage:June 5, 2013
Questions about this recall
Was Stryker Orthobiologics Aliquot Delivery System Manual Surgical Instrument - Aliquot Catheter, 6", Part Number 2110-0503 Product Usage: Facilitate placement of bone cement in weakened or diminished bone recalled?
Yes. FDA published recall Z-1399-2013 on June 5, 2013. Reported hazard: There have been a few reported cases of the catheter breaking when removing the Aliquot catheter from the needle during vertebroplasty procedures. This has been attributed to bending of the catheter during the retraction from the needle. It was determined to be use errors. The correction is limited to a revision of the Directions for Use.
What should I do if I have this product?
Voluntary: Firm initiated · Letter